Improvements to the analytical performance of inductively coupled plasma optical emission spectrometry by coupling a multi-mode sample introduction system to an infrared heated pre-evaporation tube

Author affiliations

Abstract

A multimode sample introduction system (MSIS) was coupled to a pre-evaporation tube, which was connected to the base of the torch in inductively coupled plasma optical emission spectrometry (ICPOES). The pre-evaporation tube and bottom 7 cm (for ARCOS instrument) or 3 cm (for Agilent instrument) of the torch were surrounded by an infrared-heated ceramic rope, which was heated to 80 °C in nebulisation mode and to 100 °C or 175 °C in hydride generation and dual modes. This resulted in improved sensitivity and detection limit for many elements, especially hydride-forming ones, compared to what is achieved with pneumatic nebulisation, but in some degradation of precision. The applicability of the approach was demonstrated through the accurate determination of several elements in two standard reference materials (waste water and corn bran) with matrix matching but without internal standardization. With infrared heating and operating MSIS in hydride generation (HG) mode, more elements than expected could be accurately determined, including Be, Co, Cr, Mg, Mn, etc., whose determination had not been reported in this mode before.

Graphical abstract: Improvements to the analytical performance of inductively coupled plasma optical emission spectrometry by coupling a multi-mode sample introduction system to an infrared heated pre-evaporation tube

Back to tab navigation

Supplementary files

Publication details

The article was received on 16 Feb 2018, accepted on 26 Apr 2018 and first published on 26 Apr 2018


Article type: Technical Note
DOI: 10.1039/C8JA00051D
Citation: J. Anal. At. Spectrom., 2018,33, 1068-1075
  •   Request permissions

    Improvements to the analytical performance of inductively coupled plasma optical emission spectrometry by coupling a multi-mode sample introduction system to an infrared heated pre-evaporation tube

    T. K. Anderlini and D. Beauchemin, J. Anal. At. Spectrom., 2018, 33, 1068
    DOI: 10.1039/C8JA00051D