Direct aqueous injection of the fluoroacetate anion in potable water for analysis by liquid chromatography tandem mass-spectrometry

Abstract

Sodium fluoroacetate or Compound 1080 is a rodenticide registered in the United States for use in livestock protection collars. The collars are employed to control predation on herd animals (i.e., killing of cattle by wolves or coyotes). Sodium fluoroacetate is acutely toxic to humans and has potential to cause mass casualties if used to intentionally contaminate water systems. The U.S. Environmental Protection Agency (EPA) is responsible for characterization and remediation if such an incident occurs in the civilian sector. In support of that goal, EPA has published the Selected Analytical Methods for Remediation and Recovery (SAM) document that provides sampling and analysis methods for many hazardous chemicals such as sodium fluoroacetate. Ideal SAM methods require limited sample preparation steps and utilize widely available equipment to ensure the ability for maximum laboratory participation in a large-scale response. The present paper describes a direct aqueous injection (DAI) method for liquid chromatography/tandem mass spectrometry (LC-MS/MS) analysis of the fluoroacetate anion (FAA) in potable water. Sample preservation and filtration are the only pre-processing steps required. FAA is chromatographically separated on an octylsilane (C8) reversed phase column. Separation is attributed to ion-exchange interactions. Electrospray ionization (ESI) in negative mode and detection by tandem mass spectrometry follow. FAA presence was confirmed by two fragment ions in the correct ratio, and use of a labeled standard allowed for quantitation by isotope dilution. FAA detection and quantitation limits were 0.4 μg L−1 and 2 μg L−1, respectively. Four different drinking water utilities provided water samples from varying locations across the U.S. All the water samples were fortified with FAA and tested to evaluate analyte stability and the robustness of the method.

Graphical abstract: Direct aqueous injection of the fluoroacetate anion in potable water for analysis by liquid chromatography tandem mass-spectrometry

Supplementary files

Article information

Article type
Paper
Submitted
17 Sep 2018
Accepted
27 Oct 2018
First published
29 Oct 2018

Anal. Methods, 2018,10, 5524-5531

Direct aqueous injection of the fluoroacetate anion in potable water for analysis by liquid chromatography tandem mass-spectrometry

E. Parry and S. A. Willison, Anal. Methods, 2018, 10, 5524 DOI: 10.1039/C8AY02046A

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.

Social activity

Spotlight

Advertisements