Multielemental determination in oil matrices diluted in xylene by ICP-MS with a dynamic reaction cell employing methane as reaction gas for solving specific interferences

Abstract

The analysis of organic samples by ICP-MS is still a challenge due to the occurrence of non-spectral and spectral interferences, which usually requires time-consuming sample preparation procedures for their minimization. In this work, results are presented on the direct multielemental determination in oil samples after simple dilution in xylene. The solutions were introduced into the plasma by a sample introduction system especially suited for organics, and the operational conditions of the plasma and the dynamic reaction cell were optimized. A central composite design showed that the nebulizer and the auxiliary gas flow rates had a significant influence on interference formation, monitored by Ba++ and LaO+ intensities. The compromised plasma conditions were: RF power of 1300 W and gas flow rates of 0.42 L min−1 for the nebulizer (Ar), 1.1 L min−1 for the auxiliary gas (Ar), and 0.1 L min−1 of O2 used as an additional gas. In order to minimize spectral interferences on Ca, Cr, Fe and Mg, the operational conditions of the dynamic reaction cell, working with methane as reaction gas, were optimized for 56Fe, 52Cr, 40Ca and 24Mg. Two certified reference materials, NIST 1634c (used fuel oil) and NIST 1085b (lubricating oil), were analyzed and the recoveries were in the range of 90% to 110% for most elements, being the sample detection limits between 0.2 μg kg−1 (Cr) and 40 μg kg−1 (Zn). Reliability of the data was also tested by comparing the results for a set of biodiesel samples with those determined by ICP OES and good agreement was obtained for most elements. Differences in the elemental compositions of biodiesel samples produced from different raw materials were observed, suggesting that elemental fingerprinting may be used for source identification.

Graphical abstract: Multielemental determination in oil matrices diluted in xylene by ICP-MS with a dynamic reaction cell employing methane as reaction gas for solving specific interferences

Article information

Article type
Paper
Submitted
07 Mar 2012
Accepted
21 May 2012
First published
26 Jun 2012

J. Anal. At. Spectrom., 2012,27, 1280-1286

Multielemental determination in oil matrices diluted in xylene by ICP-MS with a dynamic reaction cell employing methane as reaction gas for solving specific interferences

J. R. de Souza, C. B. Duyck, T. C. O. Fonseca and T. D. Saint'Pierre, J. Anal. At. Spectrom., 2012, 27, 1280 DOI: 10.1039/C2JA30085K

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.

Social activity

Spotlight

Advertisements