Issue 5, 2011

Fluorescence observation supporting capillary chromatography based on tube radial distribution of carrier solvents under laminar flow conditions

Abstract

We developed a capillary chromatography system by using an open capillary tube made of fused-silica, polyethylene, or polytetrafluoroethylene, and a water–hydrophilic/hydrophobic organic mixture carrier solution, called tube radial distribution chromatography (TRDC) system. By comparing with chromatograms obtained via the TRDC system, fluorescence photographs and profiles of the fluorescent dyes dissolved in the carrier solvents in capillary tubes were observed under laminar flow conditions. The chromatograms were obtained for a model mixture analyte consisting of 1-naphthol and 2,6-naphthalenedisulfonic acid with the TRDC system, by using a fused-silica capillary tube and a water–acetonitrile–ethyl acetate carrier solution. By altering the carrier flow rates, we examined the fluorescence photographs and profiles of the dyes, perylene and Eosin Y, dissolved in the carrier solvents in the capillary tube by using a fluorescence microscope equipped with a CCD camera. As confirmed by fluorescence observations, the major inner and minor outer phases generated in the capillary tube were based on the tube's radial distribution of the carrier solvents. We designed and manufactured a microreactor incorporating microchannels in which three narrow channels combined to form one wide channel. When the carrier solvents containing the dyes were fed into the channels, the inner and outer phase generations were also observed in the narrow and wide channels, strongly supporting the conclusions concerning the tube radial distribution phenomenon of the solvents.

Graphical abstract: Fluorescence observation supporting capillary chromatography based on tube radial distribution of carrier solvents under laminar flow conditions

Article information

Article type
Paper
Submitted
25 Oct 2010
Accepted
11 Nov 2010
First published
02 Dec 2010

Analyst, 2011,136, 927-932

Fluorescence observation supporting capillary chromatography based on tube radial distribution of carrier solvents under laminar flow conditions

N. Jinno, M. Murakami, K. Mizohata, M. Hashimoto and K. Tsukagoshi, Analyst, 2011, 136, 927 DOI: 10.1039/C0AN00820F

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.