Issue 7, 2010

Boron isotope ratio (δ11B) measurements in Water Framework Directive monitoring programs: comparison between double focusing sector field ICP and thermal ionization mass spectrometry

Abstract

The aim of our research was to compare δ11B measurements performed with thermal ionization mass spectrometry (TIMS) and sector field-inductively coupled plasma-mass spectrometry (SF-ICP-MS) and evaluate the feasibility of implementing stable isotope methods in European water framework directive (WFD) monitoring programs. The comparison was based on δ11B measurements of 192 ground- and surface water samples and 15 leachates of nitrate pollution source materials (organic and mineral fertilisers). The precision of δ11B measurements attainable with SF-ICP-MS, 2σ= ± 2.6‰; (n = 192), is as expected lower than the precision achieved by TIMS, 2σ= ± 0.3‰ (n = 183). However the ease of use, rapidity and availability of SF-ICP-MS on one hand and the observed variability in δ11B in ground- and surface water on the other (from −3.4 to +37‰), demonstrates that using SF-ICP-MS as an isotopic screening method would promote the use of isotopic methodology for WFD monitoring. Based on the results of the different case studies it is shown that retrieving precise information on the identification of pollution sources from δ11B values requires reaching the best analytical precision and accuracy possible. Hence, the superior precision of TIMS advantages tracing of nitrate pollution sources. However for some cases, e.g. trying to decipher contributions between sources with really distinct δ11B signatures (e.g. manure and sewage effluent), SF-ICP-MS results lead to the same conclusions and can therefore be used as a first approachable screening method for the determination of δ11B in WFD monitoring programs.

Graphical abstract: Boron isotope ratio (δ11B) measurements in Water Framework Directive monitoring programs: comparison between double focusing sector field ICP and thermal ionization mass spectrometry

Article information

Article type
Paper
Submitted
27 Jan 2010
Accepted
26 Mar 2010
First published
27 Apr 2010

J. Anal. At. Spectrom., 2010,25, 964-974

Boron isotope ratio (δ11B) measurements in Water Framework Directive monitoring programs: comparison between double focusing sector field ICP and thermal ionization mass spectrometry

K. Tirez, W. Brusten, D. Widory, E. Petelet, A. Bregnot, D. Xue, P. Boeckx and J. Bronders, J. Anal. At. Spectrom., 2010, 25, 964 DOI: 10.1039/C001840F

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.

Social activity

Spotlight

Advertisements