Issue 7, 2006

Needle enzyme electrode based glucose diffusive transport measurement in a collagen gel and validation of a simulation model

Abstract

Rapid response needle enzyme electrodes were fabricated to measure the glucose concentration at the centre of a cylindrical spiralled collagen gel, which is a relevant constituent for tissue engineering scaffolds. The experimental data were based on a low consumption glucose sensor which minimised the distorting effect of enzymatic degradation. As the measurement was carried out within a collagen gel the stirring independence was compulsory for the biosensor. Glucose concentration changes were derived from a model based on the solution to Fick's Second Law. This had two different expressions for different dimensionless time (T) domains. The expression for large T and a first order approximation for small T were known. The expression for high order approximation for small T was then derived. An analytical expression consisting of fast convergent parts of these two expressions is proposed, which operates for the entire time region. A computational model for glucose concentration evolution where an electrode is located is proposed to operate for extended time periods. The model was confirmed by agreement between the simulated and observed data. An experimental technique is developed here to determine glucose diffusion coefficient by fitting the simulated concentration profile to the observed one. The glucose diffusion coefficient within the collagen gel was estimated to be 1.3 × 10−6 cm2 s−1; higher accuracy is achieved here because errors due to noise, baseline and zero time determination are minimised with best fit.

Graphical abstract: Needle enzyme electrode based glucose diffusive transport measurement in a collagen gel and validation of a simulation model

Supplementary files

Article information

Article type
Paper
Submitted
10 Jan 2006
Accepted
30 Mar 2006
First published
16 May 2006

Analyst, 2006,131, 816-821

Needle enzyme electrode based glucose diffusive transport measurement in a collagen gel and validation of a simulation model

Z. Rong, U. Cheema and P. Vadgama, Analyst, 2006, 131, 816 DOI: 10.1039/B600334F

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.