Isotope dilution with ICP-MS—simplified uncertainty estimation using a robust procedure based on a higher target value of uncertainty

Abstract

The objective of this paper is to bring forward a simple tool for uncertainty estimation in ID-MS (isotope dilution mass spectrometry) using a robust measurement procedure. This basic form does not allow the ultimate performance to be obtained regarding uncertainty, but will still produce results that have reliable uncertainties which can be more difficult to evaluate with other calibration strategies (such as external calibration, internal standard, standard addition). The paper describes an a priori simulation process of the uncertainty that will be achieved using a validated procedure for an analyte in a matrix. For this, a valid model equation is built, where the different influence quantities are allowed to vary within a wide, robust tolerance window. For these influence quantities an approximate measurement uncertainty contribution relative to the measurand (i.e., quantity to be measured) is calculated and only those that significantly affect the measurement uncertainty (expanded, k = 2, relative) at the target level are taken into account. A target value of uncertainty of 2% is used here as an example since it is deemed to be achievable by experienced laboratories that are not primarily concerned with measuring at a lower uncertainty level. The approach is demonstrated in the determination of nickel in low-alloy steel and copper in fresh water. However, this basic approach using target uncertainty and wide tolerance windows is general and can easily be adapted for other applications.

Graphical abstract: Isotope dilution with ICP-MS—simplified uncertainty estimation using a robust procedure based on a higher target value of uncertainty

Article information

Article type
Paper
Submitted
01 Apr 2005
Accepted
09 Aug 2005
First published
23 Aug 2005

J. Anal. At. Spectrom., 2005,20, 1024-1029

Isotope dilution with ICP-MS—simplified uncertainty estimation using a robust procedure based on a higher target value of uncertainty

B. Magnusson, I. Trešl and C. Haraldsson, J. Anal. At. Spectrom., 2005, 20, 1024 DOI: 10.1039/B504576B

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.

Social activity

Spotlight

Advertisements