Abstract

A new pneumatic concentric nebulizer (flow focusing pneumatic nebulizer, FFPN), based on the capillary flow focusing effect which produces highly controllable fine aerosols, is presented for the first time for analytical applications. The key aerosol diagnostic data and analytical figures of merit obtained using ICP-AES are presented for the FFPN and compared with a commercially available glass concentric nebulizer (conical nebulizer, CK). To perform a fair comparison, both nebulizers have been run under their own standard working conditions. Under all the conditions studied, the FFPN produces a finer primary aerosol than the conventional pneumatic nebulizer. At a solution uptake rate of 2.0 mL min−1 and a nebulizer gas flow rate of 1.0 L min−1 for the CK, and at a solution uptake rate of 0.2 mL min−1 and a nebulizer gas flow rate of 0.7 L min−1 for the FFPN, the median diameter (D50) values of the primary aerosol are 24.64 μm and 3.82 μm with the CK and FFPN, respectively. The tertiary aerosols of the CK and FFPN studied do not show a substantial difference, the D50 values, under the same operating conditions stated above, being 2.89 μm and 2.87 μm for the CK and FFPN, respectively. Under most of the studied conditions, all the primary aerosols produced by FFPN are contained on droplets smaller than 10 μm. For example, under the above experimental conditions, nearly 22% and 98% of primary aerosol volume is contained on droplets having sizes less than 10 μm for the CK and FFPN, respectively. However, approximately all the tertiary aerosol volume is contained on droplets smaller than this diameter with both nebulizers. This is the result of strong filtering action of the cyclonic spray chamber used, with a cut-off diameter of 15 μm, when CK is used. In general, sensitivities, precision and limits of detection with the FFPN are similar or better than those of the CK, yet with a sample uptake rate of only one-tenth the CK rate.

Article information

Article type
Paper
Submitted
12 Jan 2004
Accepted
27 May 2004
First published
25 Aug 2004

J. Anal. At. Spectrom., 2004,19, 1340-1346

Preliminary characterization and fundamental properties of aerosols generated by a flow focusing pneumatic nebulizer

B. Almagro, A. M. Gañán-Calvo and A. Canals, J. Anal. At. Spectrom., 2004, 19, 1340 DOI: 10.1039/B400502N

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.

Social activity

Spotlight

Advertisements