In the present work the advantages and limitations of introducing the sample directly into the plasma through the use of a direct injection high-efficiency nebulizer (DIHEN) have been evaluated in ICP-AES. The reference system was a high-efficiency nebulizer (HEN), precursor of the DIHEN, coupled to a cyclonic spray chamber. It was found that the plasma can only be operated under non-robust conditions because of the high solvent loading. The issues considered in the present work were: (i) the aerosol characteristics; (ii) the ICP-AES analytical figures of merit; and (iii) the interferences induced by the presence of inorganic as well as organic matrices. The results indicated that the aerosols obtained by the DIHEN are coarser than those leaving the cyclonic spray chamber. Nonetheless, and due to the higher analyte transport efficiency, the DIHEN affords ICP-AES sensitivities and LODs up to 10–18 times better than the cyclonic spray chamber. Also, it was found that, under the plasma conditions tested, the optimum nebulizer gas and liquid flow rates in terms of sensitivity for the DIHEN were very different from those for the HEN with the cyclonic spray chamber. In the first case, the signal peaked at rather low liquid (Ql) and gas (Qg) flow rates, whereas for the cyclonic spray chamber moderate and/or high liquid and gas flow rates were advisable. Under a given set of plasma conditions, the DIHEN has been proven to produce problems due to deterioration in the plasma, derived from the higher solvent load with respect to the HEN coupled to the cyclonic spray chamber. Despite these findings and the fact that a non-robust plasma was used, the matrix effects for inorganic and organic species were significantly minimized, for a low Ql and for some lines, in the case of the DIHEN.

Article information

Article type
Paper
Submitted
24 Nov 2000
Accepted
22 Feb 2001
First published
28 Mar 2001

J. Anal. At. Spectrom., 2001,16, 514-520

Evaluation of a direct injection high-efficiency nebulizer (DIHEN) by comparison with a high-efficiency nebulizer (HEN) coupled to a cyclonic spray chamber as a liquid sample introduction system for ICP-AES

J. Todolí and J. Mermet, J. Anal. At. Spectrom., 2001, 16, 514 DOI: 10.1039/B009430G

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.

Social activity

Spotlight

Advertisements