Issue 12, 2000

Abstract

A sensitive and robust method for the determination of eight inorganic and organic arsenic species by ion-exclusion liquid chromatography combined with inductively coupled plasma mass spectrometry (LC-ICP-MS) and LC-hydride generation-ICP-MS (LC-HG-ICP-MS) is described. The species are arsenite (AsIII), arsenate (AsV), monomethylarsonic acid (MMAs), dimethylarsinic acid (DMAs), arsenobetaine (AsB), trimethylarsine oxide (TMAsO), tetramethylarsonium salt (TMAs) and arsenocholine salt (AsC). A good separation of seven arsenic species, AsIII, AsV, MMAs, DMAs, AsB, TMAsO, and AsC or TMAs, was achieved by using an ion-exclusion column packed with a carboxylated methacrylate resin and 0.35 mmol l−1 of a sodium sulfate solution adjusted to pH 3.8 as the mobile phase. The detection limits of the eight arsenic species obtained by LC-ICP-MS ranged from 0.067 to 0.34 ng As ml−1 using an injection volume of 50 µl. A hydride generation technique improved the detection limits of AsIII, AsV, MMAs, DMAs and TMAsO to 0.016–0.075 ng As ml−1, while AsB, TMAs and AsC were not detectable. The relative standard deviations of five replicates of a standard of each arsenic species by the former method ranged from 1.3 to 3.3% and those by the latter method, from 1.8 to 3.1%. The proposed methods were successfully applied to the determination of five arsenic species in human urine and six species in an extract from tuna fish tissue. The only necessary pretreatment for these analyses involved filtration with a 0.45 µm membrane and the ArCl polyatomic interference, due to a large amount of chloride in the biological samples on 75As measurement, was eliminated by the described LC separation. In addition, serious deterioration in column performance and a decrease in the sensitivity of ICP-MS were not observed during the experimental period of five months. The LC-ICP-MS and LC-HG-ICP-MS methods were validated by analyzing reference samples of human urine and tuna fish tissue.

Article information

Article type
Paper
Submitted
25 Jul 2000
Accepted
02 Oct 2000
First published
01 Nov 2000

J. Anal. At. Spectrom., 2000,15, 1546-1552

Ion-exclusion chromatography combined with ICP-MS and hydride generation-ICP-MS for the determination of arsenic species in biological matrices

T. Nakazato, T. Taniguchi, H. Tao, M. Tominaga and A. Miyazaki, J. Anal. At. Spectrom., 2000, 15, 1546 DOI: 10.1039/B005981L

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.

Social activity

Spotlight

Advertisements