Determination of arsenic in biological fluids by electrothermal atomic absorption spectrometry

Abstract

A procedure for the determination of the total content of arsenic in urine, serum and blood by electrothermal atomic absorption spectrometry (ETAAS) is described. Zeeman correction is used to compensate the high background signals. The samples are diluted (1 + 1 for urine and 1 + 3 for both serum and blood samples) in a medium containing 0.1% w/v Triton X-100 before being introduced directly into the furnace. A solution containing 15% w/v hydrogen peroxide, 0.65% w/v nitric acid and 0.5% w/v nickel is also introduced into the atomizer by means of a separate injection. Calibration is carried out against aqueous standards for blood and serum samples and using the standard additions method for urine samples. The detection limit is 20 pg (2 ng ml−1). The reliability of the procedure is checked by analyzing three certified reference materials and by recovery studies.

Article information

Article type
Paper
Submitted
20 Sep 1999
First published
26 Jan 2000

Analyst, 2000,125, 313-316

Determination of arsenic in biological fluids by electrothermal atomic absorption spectrometry

N. Campillo, P. Viñas, I. López-García and M. Hernández-Córdoba, Analyst, 2000, 125, 313 DOI: 10.1039/A907596H

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.

Social activity

Spotlight

Advertisements