Abstract

A method for the automated determination of aluminum in a variety of beverages is described. The method utilizes lumogallion as a complexing agent in a buffer solution. The system is very similar to flow-injection analysis (FIA), however, the tubing id is larger than that typically used in FIA. Therefore, the system is best described as non-segmented continuous flow analysis using fluorescence spectroscopy detection. The method is extremely simple, requiring virtually no sample preparation and only one reagent. The instrument detection limit for aluminum is 0.012 µg ml–1 and calibration is linear to 3 µg ml–1. Results from a variety of beverage matrices are discussed and compared with the frequently used 8-hydroxyquinolone method utilizing a chloroform extraction and fluorescence spectroscopy detection.

Article information

Article type
Paper

Analyst, 1999,124, 793-796

Determination of aluminum in beverages by automated non-segmented continuous flow analysis with fluorescent detection of the lumogallion complex

R. L. Hoch, Analyst, 1999, 124, 793 DOI: 10.1039/A809247H

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.

Social activity

Spotlight

Advertisements