Separation-optimized sequential injection method for rapid automated analytical separation of 90Sr in nuclear waste

Abstract

An automated sequential injection (SI) analyzer was developed for analytical separations of 90Sr in nuclear waste samples. The system provides rapid separation of 90Sr from matrix and potential interfering elements using an Sr-selective extraction chromatographic column. After separation, 90Sr activity can be determined by either on-line radiometric detection or off-line counting of collected fractions. The separation was investigated under automated conditions. Eluent compositions were determined that provide consistent, high, reproducible recoveries of Sr from the separation column. Procedures were designed with emphasis on rapid sample throughput. Using fraction collection, a set of eight samples can be processed in just over 1 h. Using on-line detection, each sample is separated and quantified in 22 min. The method was applied to nuclear waste supernate and vitrified nuclear waste.

Article information

Article type
Paper

Analyst, 1999,124, 203-210

Separation-optimized sequential injection method for rapid automated analytical separation of 90Sr in nuclear waste

J. W. Grate, O. Egorov and S. K. Fadeff, Analyst, 1999, 124, 203 DOI: 10.1039/A807468B

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.

Social activity

Spotlight

Advertisements