Issue 1, 1997

Error Analysis of the Transmittance Ratio Stray Radiant Energy Test Method in Double-beam Ratio Recording Spectrophotometers

Abstract

The photometric error contributions which the relative errors in determining both the cell pathlength ratio and the transmittance ratio minimum make towards the relative error in determining the relative stray radiant energy (SRE) levels in a ratio recording double-beam spectrophotometer were investigated, thus enabling optimum experimental conditions for SRE determination to be pre-selected.

Article information

Article type
Paper

Analyst, 1997,122, 39-40

Error Analysis of the Transmittance Ratio Stray Radiant Energy Test Method in Double-beam Ratio Recording Spectrophotometers

P. Fleming and A. Fleming, Analyst, 1997, 122, 39 DOI: 10.1039/A605825F

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.