Issue 7, 1996

Separation of ifosfamide and its degradation products using micellar electrokinetic chromatography

Abstract

Ifosfamide and its thermal degradation products were separated using micellar electrokinetic chromatography (MEKC). The method was compared with a published reversed phase HPLC method. Much better resolution was obtained using the MEKC method. Nine degradation products could be detected using MEKC compared with only two using reversed phase HPLC. The LOD using the equipment described was 0.05 mg ml–1 with UV detection at 185 mm. The standard error of the peak area for five injections was 2.8%. The standard error of peak position of the degradation products varied from less than 1% up to 5% with the latter eluting peaks having the higher variation.

Article information

Article type
Paper

Anal. Commun., 1996,33, 235-237

Separation of ifosfamide and its degradation products using micellar electrokinetic chromatography

S. Hill, J. Aexiou, P. E. Kavanagh, P. C. Livingstone, R. A. Shalliker and J. Soscic, Anal. Commun., 1996, 33, 235 DOI: 10.1039/AC9963300235

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.