Re-evaluation of the thiocyanate dosimeter for pulse radiolysis
Abstract
The thiocyanate dosimeter (10–2 mol dm–3 SCN– in O2-saturated water) has been standardised against the super-Fricke dosimeter (10–2 mol dm–3 FeII in O2-saturated 0.4 mol dm–3 H2SO4) using the hexacyanoferrate(II) dosimeter [5 × 10–3 mol dm–3 Fe(CN)64– in O2-saturated water] as a secondary standard. On the basis that G(FeIII)= 1.67 × 10–6 mol J–1 and εFeIII= 220.4 m2 mol–1 at 304 nm and 25 °C in the super-Fricke dosimeter, we obtain Gε[Fe(CN)63–]=(3.47 ± 0.06)× 10–5 m2 J–1 at 420 nm and Gε(SCN)2˙–=(2.59 ± 0.05)× 10–4 m2 J–1 at 475 nm. These values remain unchanged when the solutions are saturated with air instead of O2 and are doubled in N2O-saturated solution.