Evaluation of ractopamine cross-reactivity in several commercially available β-agonist enzyme immunoassay kits
Abstract
Several enzyme immunoassay test kits are commercially available for screening of matrices including cattle feed, tissues, faeces and urine for the presence of β-agonists. The kits utilized for this evaluation offer sensitivity as low as 0.1 ng ml–1(assay solution concentration, not test sample). Evaluation of ractopamine hydrochloride (LY31537) solutions at concentrations as high as 1000 ng ml–1, using six different kits from four different manufacturers, showed cross-reactivity of less than 0.5%.