Influence of concentration and sample volume on the recovery of compounds from water following direct sorption on Tenax TA–thermal desorption

Abstract

Thermal desorption of organic compounds following their direct sorption on Tenax TA from water is described. A simple and fast method of cartridge drying after sorption is presented. The optimum drying conditions proposed were a temperature of 80 °C and a flow rate of 60 cm3 min–1. The recoveries determined for sample volumes of 5, 10 and 100 cm3 and analyte masses on the sorbent bed of 2–100 ng of toluene, chlorobenzene and m-xylene were found to be 63, 73 and 78%, respectively. The results show that recoveries do not depend on the concentrations and volumes of the investigated samples.

Article information

Article type
Paper

Analyst, 1995,120, 2781-2786

Influence of concentration and sample volume on the recovery of compounds from water following direct sorption on Tenax TA–thermal desorption

L. Wolska, W. Janicki and J. Namieśnik, Analyst, 1995, 120, 2781 DOI: 10.1039/AN9952002781

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.

Social activity

Spotlight

Advertisements