Direct determination of dissolved uranyl(VI) in sea-water by cathodic stripping voltammetry

Abstract

The effect of the adsorption of uranyl species on the static mercury drop electrode is used for the electrochemical determination of dissolved uranium concentration in untreated and estuarine sea-water. The measurements were performed by square-wave cathodic stripping voltammetry. The standard additions method is successfully applied for measurements of natural uranyl concentrations in water samples which were found to be in the range between 4.3 and 15 nmol l–1 depending on the sampling location with detection limits of 2.4 and 10 nmol l–1, respectively.

Article information

Article type
Paper

Analyst, 1995,120, 1989-1993

Direct determination of dissolved uranyl(VI) in sea-water by cathodic stripping voltammetry

R. Djogić and M. Branica, Analyst, 1995, 120, 1989 DOI: 10.1039/AN9952001989

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.