Programmed two-stage flow controller for supercritical fluid chromatography

Abstract

The flow rate in a capillary supercritical fluid chromatographic (SFC) column can be maintained near the optimum value of a Van Deemter plot throughout the entire pressure or density programme by use of a programmed two-stage flow controller. The second stage is a conventional linear restrictor, which is designed to provide less restriction than is customary in capillary SFC. The first stage consists of a length of deactivated fused silica of 7 µm i.d. held in a temperature-controlled environment. As the pressure in the column is increased during a pressure or density programme, the temperature of the 7 µm fused-silica tube can be varied to maintain a constant flow rate.

Article information

Article type
Paper

Analyst, 1994,119, 1315-1318

Programmed two-stage flow controller for supercritical fluid chromatography

D. Pyo, Analyst, 1994, 119, 1315 DOI: 10.1039/AN9941901315

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.