Issue 4, 1993

Signal enhancement of lead and thallium in inductively coupled plasma atomic emission spectrometry using on-line anodic stripping voltammetry

Abstract

An on-line anodic stripping voltammetric (ASV) flow cell was employed for the preconcentration and signal enhancement of transition metals, which exhibit relatively high detection limits in inductively coupled plasma atomic emission spectrometry but which are also responsive in ASV. Signal enhancements of over 60-fold for 20 ml samples of thallium(I) and lead(II) were obtained in much less time than enhancements previously reported for cadmium using this method. The precision was acceptable and plots of peak heights and areas versus mass of analyte were linear for thallium and lead. Plots of enhancement factor versus sample volume were also linear for both metals.

Article information

Article type
Paper

J. Anal. At. Spectrom., 1993,8, 545-550

Signal enhancement of lead and thallium in inductively coupled plasma atomic emission spectrometry using on-line anodic stripping voltammetry

J. R. Pretty and J. A. Caruso, J. Anal. At. Spectrom., 1993, 8, 545 DOI: 10.1039/JA9930800545

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.