Issue 3, 1993

Determination of manganese in river and sea-water samples by electrothermal atomic absorption spectrometry with a tungsten atomizer

Abstract

A method for the direct determination of manganese in river and sea-water reference materials using electrothermal atomization atomic absorption spectrometry with a tungsten atomizer is described. Interferences caused by the sea-water matrix are investigated and it is shown that these can be eliminated by the addition of ascorbic acid. The method is applied to the determination of manganese in sea-water reference materials and results within one standard deviation of certified values are obtained. The detection limit (3σ) in sea-water is 1.2 pg, corresponding to 0.12 µg l–1 for a 10 µl sample.

Article information

Article type
Paper

J. Anal. At. Spectrom., 1993,8, 409-413

Determination of manganese in river and sea-water samples by electrothermal atomic absorption spectrometry with a tungsten atomizer

S. Xiao-quan, B. Radziuk, B. Welz and O. Vyskočilová, J. Anal. At. Spectrom., 1993, 8, 409 DOI: 10.1039/JA9930800409

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.

Social activity

Spotlight

Advertisements