Issue 10, 1993

Determination of phosgene (carbonyl chloride) in air by high-performance liquid chromatography with a dual selective detection system

Abstract

The development of a sensitive and highly selective method for determining airborne phosgene in the workplace is described. Phosgene was collected in an impinger solution containing tryptamine, as the derivatizing reagent. The derivative was determined by a reversed-phase high-performance liquid chromatograph equipped with dual detectors, fluorescence emission and amperometric oxidation. Limits of detection for both detectors were 8 ng ml–1. As both detectors are selective, the confidence level of the analysis is considerably enhanced. Simulated air sampling for phosgene vapours generated in a glass gas-collecting bulb showed that recoveries were excellent and that the method can be used to detect phosgene at a level of 0.04 mg m–3[a tenth of the threshold limit value–time weighted average (TLV–TWA)] in a 1 | air sample.

Article information

Article type
Paper

Analyst, 1993,118, 1285-1287

Determination of phosgene (carbonyl chloride) in air by high-performance liquid chromatography with a dual selective detection system

W. S. Wu and V. S. Gaind, Analyst, 1993, 118, 1285 DOI: 10.1039/AN9931801285

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.

Social activity

Spotlight

Advertisements