Issue 11, 1992

On-line microwave sample pre-treatment for hydride generation and cold vapour atomic absorption spectrometry. Part 2. Chemistry and applications

Abstract

A system for on-line treatment of liquid samples in a microwave oven digestor was evaluated for use with cold vapour (CV) and hydride generation (HG) atomic absorption spectrometry (AAS). Various oxidation mixtures were tested and those based on bromination (bromate–bromide–acid) and peroxodisulfate (persulfate–acid–complexing agent) were found to be compatible with and most appropriate for CVAAS and HGAAS. The composition of reagents and the analytical conditions were optimized for the determination of mercury, arsenic, bismuth, lead and tin in urine and environmental waters. The limits of detection were 0.01 and 0.2 µg l–1 for mercury, with and without amalgamation, respectively, 0.5 µg l–1 for arsenic, 0.07 µg l–1 for bismuth and 0.1 µg l–1 for tin and the sample throughput was between 13 and 30 h–1.

Article information

Article type
Paper

Analyst, 1992,117, 1735-1741

On-line microwave sample pre-treatment for hydride generation and cold vapour atomic absorption spectrometry. Part 2. Chemistry and applications

D. L. Tsalev, M. Sperling and B. Welz, Analyst, 1992, 117, 1735 DOI: 10.1039/AN9921701735

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.