Determination of nickel in biological materials after microwave dissolution using inductively coupled plasma atomic emission spectrometry with prior extraction into butan-1-ol
Abstract
A sensitive procedure has been developed for the determination of ultratrace amounts of nickel in biological materials by inductively coupled plasma atomic emission spectrometry after extraction of the nickel ion into butan-1-ol by using 1,5-bis(di-2-pyridylmethylene)thiocarbonohydrazide as the extracting reagent. Fast, efficient and complete sample digestion is achieved by an HNO3—HCl poly(tetrafluoroethylene) bomb dissolution technique using microwave heating. Results obtained for eleven certified reference materials agreed with the certified values.