Analytical performance evaluation of a 40.68 MHz inductively coupled plasma mass spectrometer

Abstract

The instrumentation of a new 40.68 MHz inductively coupled plasma (ICP) mass spectrometer is described. An optimization study has been carried out for the instrument. Of the many operating parameters, the carrier argon flow rate is the most sensitive and the optimum carrier argon flow rate is found to be much lower than the literature values for a 27.12 MHz ICP mass spectrometer. The analytical performance of the instrument, i.e., the detection limits, background mass spectrum and formation of oxides and doubly charged ions is also evaluated.

Article information

Article type
Paper

J. Anal. At. Spectrom., 1991,6, 431-437

Analytical performance evaluation of a 40.68 MHz inductively coupled plasma mass spectrometer

C. J. Park and K. W. Lee, J. Anal. At. Spectrom., 1991, 6, 431 DOI: 10.1039/JA9910600431

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.