Sample introduction system for direct current plasma atomic emission spectrometry

Abstract

A dual pump-nebulisation sample introduction system for direct current plasma atomic emission spectrometry (DCP-AES) was designed both to enhance sensitivity and to reduce interferences. A solution of the sample and a solution of easily ionised element (EIE) of constant concentration were introduced separately into the DCP through two pump-nebulisation assemblies and a modified sample introduction tube. Hence the buffering effect of the EIEs was easily achieved without adding EIEs to the sample solution during sample preparation. This system was evaluated by its application to the determination of V, Cr, Mn, Fe and Ni. Ionisation interferences from the sample matrix were reduced to negligible levels. Detection limits for the elements determined were better than those obtained using a commercial DCP-AES system without modification.

Article information

Article type
Paper

J. Anal. At. Spectrom., 1990,5, 559-562

Sample introduction system for direct current plasma atomic emission spectrometry

I. D. Brindle and X. Le, J. Anal. At. Spectrom., 1990, 5, 559 DOI: 10.1039/JA9900500559

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.

Social activity

Spotlight

Advertisements