Aerosol particle size effects on plasma spectrometric analysis

Abstract

A low-flow cascade impactor was used to evaluate the effects of tertiary aerosol particle size on inductively coupled plasma atomic emission spectrometry (ICP-AES) measurements. The impactor was used either on-line as a size-selective filter during measurements, or off-line for the determination of tertiary analyte mass particle size distributions. A pneumatic nebuliser and various thermospray sample introduction systems were studied. Particle size had little effect on background noise, although a 20% drop at small cut-off diameters was observed for the thermospray and improvements in signal to background noise ratios were observed using a 6-µm cut-off. Particle size also had little effect on measurement precision, however, it was important in determining how much signal was produced per unit mass of analyte transported. Analyte residing in aerosol particles of ⩽2.5 µm was shown to be much more efficiently utilised for signal production than that in larger droplets.

Article information

Article type
Paper

J. Anal. At. Spectrom., 1989,4, 291-294

Aerosol particle size effects on plasma spectrometric analysis

J. A. Koropchak and H. Aryamanya-Mugisha, J. Anal. At. Spectrom., 1989, 4, 291 DOI: 10.1039/JA9890400291

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.

Social activity

Spotlight

Advertisements