Simulation of flow injection transients with application to amperometric detector response

Abstract

An RC (resistance-capacitor) circuit has been built as an analogue simulator of the continuously stirred tank reactor model of the flow injection transport process. The analogue simulator was used as a fast-response time-calibration signal and to evaluate the fidelity of the time response of amperometric detection systems. Further uses of the device are discussed.

Article information

Article type
Paper

Analyst, 1989,114, 799-803

Simulation of flow injection transients with application to amperometric detector response

F. E. Powell and A. G. Fogg, Analyst, 1989, 114, 799 DOI: 10.1039/AN9891400799

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.