Accurate measurement of stable isotopes of magnesium in biological materials with inductively coupled plasma mass spectrometry

Abstract

A method is described for the accurate isotopic determination of magnesium (24Mg, 25Mg, 26Mg) in biological materials, which is based on inductively coupled plasma mass spectrometry (ICP-MS). The analytical performance of the method was examined with respect to the requirements of stable isotope tracer studies. When applied to the measurement of base-line isotope ratios (MRa/b) in rat tissue the following results were obtained (data are given in terms of MR25/24 and MR26/24± 1RSD, n= 4 or 5): standard solution of Mg (0.05 µg ml–1), 0.1394 ± 0.6, 0.1697 ± 0.8; bone, 0.1401 ± 0.5, 0.1706 ± 0.8; brain, 0.1404 ± 0.2, 0.1720 ± 0.3; kidney, 0.1392 ± 0.5, 0.1702 ± 0.6; liver, 0.1388 ± 0.4, 0.1696 ± 0.5; muscle, 0.1396 ± 0.2, 0.1716 ± 0.4; plasma, 0.1385 ± 0.3, 0.1691 ± 0.3; red cells, 0.1383 ± 0.1, 0.1694 ± 0.2; and urine, 0.1402 ± 0.1, 0.1721 ± 0.3. The measurement precision for replicate analyses of each matrix was in the range 0.1–1.0% and the mean value of the isotope ratio for different matrices agreed with the corresponding ratio for the standard solution to within 1.5%. When measured over a 10-h period, the isotope ratios appeared to be independent of the observed drifts in the ion beam intensities. The instrument blank contribution to the ion beam intensities was about 0.1% of the values obtained for a solution containing 0.05 µg ml–1 of natural Mg. The absolute detection limit for Mg (based on the experimental standard deviation of blanks run over 10 h) was <2 ng for all three stable isotopes. The ICP-MS phase of the analysis was capable of making 50–150 separate measurements of both ratios in a 10-h period, depending on the desired level of measurement precision within the range 0.1–1.0%. A complete analytical scheme for the accurate determination of the three stable isotopes of Mg, which is based on precipitation with ammonium phosphate, is described. The accuracy of the method was tested using the standard reference materials Bovine Liver (NBS 1577a, 600 ± 15 µg g–1 of Mg) and Animal Bone (IAEA H-5, 3550 ± 90 µg g–1). The proposed method provided the following data (µg g–1): SRM 1577a, 617 ± 4 and IAEA H-5, 3585 ± 16. For other biological matrices of interest, the accuracy of the method was compared with atomic absorption spectrometry. The complete analytical procedure, up to the point of mass spectrometric measurement, can be performed on about 20 samples per working day.

Article information

Article type
Paper

Analyst, 1988,113, 1837-1842

Accurate measurement of stable isotopes of magnesium in biological materials with inductively coupled plasma mass spectrometry

S. Schuette, D. Vereault, B. T. G. Ting and M. Janghorbani, Analyst, 1988, 113, 1837 DOI: 10.1039/AN9881301837

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.

Social activity

Spotlight

Advertisements