A continuous flow system was constructed for the determination of sulphate in the concentration range 10–5–10–3M. The reagent stream, containing lead(II) ion, ionic-strength adjuster and pH-adjustment buffer in methanol, was mixed with a sample sulphate stream and the free lead(II) concentration was monitored with a lead ion-selective electrode. The calibration graph is not linear owing to the titrimetric character of the determination and the curvature is explained by the solubility equilibrium that exists between lead and sulphate. The relative error was within ±10%. Interference from hydrogen carbonate was observed even in the presence of acidic buffer.

Article information

Article type
Paper

Analyst, 1988,113, 1817-1820

Continuous flow determination of sulphate with a lead-selective electrode

H. Hara, G. Horvai and E. Pungor, Analyst, 1988, 113, 1817 DOI: 10.1039/AN9881301817

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.