Issue 2, 1987

Correction of matrix effects in inductively coupled plasma atomic emission spectrometry by interactive power adjustment

Abstract

An important class of matrix effect in inductively coupled plasma atomic emission spectrometry is apparently caused by matrix-induced changes in excitation conditions in the plasma, as spectral lines are affected to an extent that depends on their excitation potentials. A change in the r.f. power supplied to the plasma produces a similar effect. This similarity has been used to compensate for matrix effects by means of an automatic increase in forward power. Excitation conditions were monitored by the sensitivity of a reference line. Matrix effects on most of the lines studied were reduced to a small level by the method, which is termed “interactive power adjustment.”

Article information

Article type
Paper

J. Anal. At. Spectrom., 1987,2, 185-188

Correction of matrix effects in inductively coupled plasma atomic emission spectrometry by interactive power adjustment

M. Thompson, M. H. Ramsey, B. J. Coles and C. M. Du, J. Anal. At. Spectrom., 1987, 2, 185 DOI: 10.1039/JA9870200185

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.