Improvement of a pneumatic nebuliser for atomic absorption spectrometry

Abstract

The importance of the liquid transport characteristics of atomic absorption nebulisers is discussed and a simple method for determining these characteristics reported. This method was used to optimise the pumping characteristics of a commercial pneumatic nebuliser. Compared with the original nebuliser, the optimised version is significantly less sensitive to variations in the level of sample liquid (approximately a factor of five improvement). The effects of the nebulising gas pressure and sample uptake rate are presented and the improved analytical performance of the new nebuliser is shown.

Article information

Article type
Paper

J. Anal. At. Spectrom., 1986,1, 55-58

Improvement of a pneumatic nebuliser for atomic absorption spectrometry

B. T. Sturman, J. Anal. At. Spectrom., 1986, 1, 55 DOI: 10.1039/JA9860100055

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.