Abstract

A spectrophotometric method with flow injection analysis is described for the determination of vitamin C in urine, which presents substantial improvements over the existing conventional manual and automatic methods. Determinations can be carried out at the µg ml–1 level; the method has an average recovery error of ±2.5% and a sampling frequency of 90 samples h–1 can be attained.

Article information

Article type
Paper

Analyst, 1986,111, 167-169

Determination of vitamin C in urine by flow injection analysis

F. Lázaro, A. Ríos, M. D. L. de Castro and M. Valcárcel, Analyst, 1986, 111, 167 DOI: 10.1039/AN9861100167

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.