Abstract

Flow injection analysis offers many advantages as a sample introduction technique for atomic-absorption spectroscopy. The methods of sample injection, however, are wasteful of sample. Controlled-dispersion analysis is a technique that dispenses with the need for injection septa or valves and consumes a minimum amount of sample.

Methods are described for the determination of lithium, magnesium, calcium, copper and zinc in body fluids. As the sample volume can be easily varied over a wide range, calibration can be performed using only one standard. The methods are precise and easily automated and have a sample throughput greater than 120 samples h–1. A unique method of economising on reagent is also described.

Article information

Article type
Paper

Analyst, 1985,110, 493-496

Controlled-dispersion flow analysis with atomic-absorption detection for the determination of clinically relevant elements

R. A. Sherwood, B. F. Rocks and C. Riley, Analyst, 1985, 110, 493 DOI: 10.1039/AN9851000493

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.

Social activity

Spotlight

Advertisements