Limit of detection in barium sulphate gravimetry for water analysis

Abstract

Sulphate ions in the 1–5 mg l–1 SO42– range can be determined gravimetrically in 200-ml samples using a large excess of barium chloride. From ten determinations at each of three concentration levels of standard sodium sulphate solution, the analytical calibration function calculated by weighted least squares is xc(mg l–1 SO42– found)=(0.093 ± 0.064)+(1.008 ± 0.039)c(mg l–1 SO42– added) with 95% confidence intervals. The limit of detection is cL= 1 mg l–1 SO42–(k= 3). A systematic approach is suggested for determining the limit of detection, by considering the variance of both the blank and the calibration function.

Article information

Article type
Paper

Analyst, 1985,110, 403-406

Limit of detection in barium sulphate gravimetry for water analysis

R. Ferrús and F. Torrades, Analyst, 1985, 110, 403 DOI: 10.1039/AN9851000403

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.