Issue 8, 1984

Simple and sensitive ion chromatograph for trace metal determination

Abstract

A simple non-suppressed ion-chromatographic system is described for the determination of cations. The design of a novel detector and post-column reactor, exhibiting high sensitivity and low dispersion, are described in detail together with an on-column cation concentration procedure. The system has a linearity of over three orders of magnitude and is satisfactory for determining cations present in aqueous media down to a level of a few parts per billion in the presence of high concentrations of salt and organic materials.

Article information

Article type
Paper

Analyst, 1984,109, 997-1002

Simple and sensitive ion chromatograph for trace metal determination

G. J. Schmidt and R. P. W. Scott, Analyst, 1984, 109, 997 DOI: 10.1039/AN9840900997

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.