Issue 1276, 1982

Determination of trace impurities in cumene by capillary gas chromatography

Abstract

Gas-chromatographic conditions have been established for the determination of trace hydrocarbons in cumene feed, employing a capillary (WCOT) column of UCON LB-550X under isothermal conditions. Retention indices for aromatic hydrocarbons are reported at five temperatures (90, 100, 110, 120 and 130 °C). Decane was used as an internal standard and precision data are reported for quantitative calculations. Under the reported conditions, it is possible to determine 5 p.p.m. of impurity without any pre-concentration in less than 30 min.

Article information

Article type
Paper

Analyst, 1982,107, 761-769

Determination of trace impurities in cumene by capillary gas chromatography

B. Kumar, H. S. Mathur, R. K. Kuchhal and P. L. Gupta, Analyst, 1982, 107, 761 DOI: 10.1039/AN9820700761

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.