Issue 1272, 1982

Optical emission spectrometry with an inductively coupled radiofrequency argon plasma source and direct sample introduction from a graphite rod

Abstract

Preliminary studies of analytical performance are described for an instrumental assembly in which microlitre volumes of liquid samples are applied to a graphite rod, desolvated and the rod is inserted axially directly into a continuously operating low-power inductively coupled argon plasma. Simplex optimisation of operating parameters has been undertaken for manganese to determine the most favourable signal to background intensity ratios and detection limits are reported for manganese, cadmium, cobalt, copper, iron, nickel and lead.

Article information

Article type
Paper

Analyst, 1982,107, 276-281

Optical emission spectrometry with an inductively coupled radiofrequency argon plasma source and direct sample introduction from a graphite rod

G. F. Kirkbright and S. J. Walton, Analyst, 1982, 107, 276 DOI: 10.1039/AN9820700276

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.