Automatic determination of orthophosphate, ammoniacal nitrogen and total carbohydrate in fermentation media

Abstract

The automatic simultaneous determination of orthophosphate, ammoniacal nitrogen and total carbohydrate in aqueous samples from fermentation processes is described. Optimum conditions were established that prevented interferences between methods and catered for samples of different pH values and carbohydrate concentrations.

The capacity of the AutoAnalyzer is twenty samples per hour, with a minimum of sample handling and pre-treatment. Relative standard deviations obtained were in the range 1·0 to 3·2 per cent., a significant improvement on manual procedures. A statistical comparison of results indicated that there was no bias between the automatic and manual procedures.

Article information

Article type
Paper

Analyst, 1970,95, 738-742

Automatic determination of orthophosphate, ammoniacal nitrogen and total carbohydrate in fermentation media

C. Maddix, R. L. Norton and N. J. Nicolson, Analyst, 1970, 95, 738 DOI: 10.1039/AN9709500738

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.

Social activity

Spotlight

Advertisements