Marco
Girolami
*a,
Matteo
Bosi
b,
Valerio
Serpente
a,
Matteo
Mastellone
a,
Luca
Seravalli
b,
Sara
Pettinato
c,
Stefano
Salvatori
c,
Daniele M.
Trucchi
a and
Roberto
Fornari
db
aIstituto di Struttura della Materia, Consiglio Nazionale delle Ricerche (ISM – CNR), Sede Secondaria di Montelibretti, Strada Provinciale 35D, 9, 00010 Montelibretti, Roma, Italy. E-mail: marco.girolami@ism.cnr.it
bIstituto dei Materiali per l’Elettronica e il Magnetismo, Consiglio Nazionale delle Ricerche (IMEM – CNR), Parco Area delle Scienze 37/A, 43124, Parma, Italy
cFaculty of Engineering, Università degli Studi Niccolò Cusano, Via Don Carlo Gnocchi 3, 00166 Roma, Italy
dDepartment of Mathematical, Physical and Computer Sciences, Università di Parma, Parco Area delle Scienze 7/A, 43124 Parma, Italy
First published on 28th February 2023
Photoelectronic properties of orthorhombic undoped κ-Ga2O3 epitaxial thin films, grown on sapphire substrates by metal–organic vapour phase epitaxy, were evaluated under X-ray irradiation (CuKα line, 8.05 keV) for the first time. Photoresponse linearity at low dose-rates (varying in the 10–200 μGy s−1 range), and excellent detection sensitivity (up to 342.3 μC Gy−1 cm−3), were demonstrated even at very low applied electric fields (down to 0.001 V μm−1). Photocurrent rise time was evaluated to be <0.5 s, and signal stability was assessed for exposure times up to 2 h, highlighting no degradation of the performance. These encouraging results, mostly due to the extremely low dark current measured (in the pA range), suggest that orthorhombic undoped κ-Ga2O3 is a promising material for the fabrication of sensitive and stable large-area X-ray detectors with minimum power consumption.
In this context, gallium oxide (Ga2O3), one of the most promising wide-bandgap semiconductors for high power electronics and solar-blind UV photodetectors, can be considered as an attractive option, because it offers a convenient trade-off between different materials investigated as X-ray photoconductors: in particular, the combination of a wide bandgap (in the 4.7–5.2 eV range, depending on the considered polymorph), ensuring a very low dark current, with a high density (6.4 g cm−3), providing efficient X-ray absorption, makes Ga2O3 a sensitive and stable active material for X-ray detectors even at low applied electric fields.12 Moreover, it is radiation-hard13 and it can be deposited over large areas.14
Ga2O3 is known since about seven decades, but only recently it has attracted renewed attention thanks to significant technological advancements: (1) high-quality single crystals (up to 4” diameter) can now be achieved using both edge-defined film-fed (EFG)15 and Czochralski16 growth techniques; (2) advanced epitaxial techniques have been developed, leading to good quality homo- and hetero-epitaxial films with reproducible properties; and (3) electrical properties can be tailored by appropriate doping techniques, providing conductivity ranging from high n-type to semi-insulating. A review of the most recent advances in the field of Ga2O3 can be found in ref. 17.
Five Ga2O3 polymorphs are known, i.e., α, β, γ, δ, and κ, but research has been mostly concentrated on the monoclinic (β) one, as it is thermodynamically stable, it can be directly grown from the melt, and can withstand the high-temperature treatments necessary for the activation of the implanted dopants. However, in the past few years, Ga2O3 metastable polymorphs have attracted attention owing to their peculiar properties. For instance, the corundum (α) phase exhibits hexagonal symmetry and the widest bandgap (5.2 eV), making it attractive for power devices.18 The orthorhombic (κ) phase can be easily deposited by MOVPE (metal–organic vapour phase epitaxy) on commercial (0001)-oriented sapphire14 substrates as nominally undoped high-resistivity epilayers, which can conveniently be exploited in solar blind UV-C photodetectors.19 It is worth noting here that often the literature indifferently reports κ- and ε-Ga2O3; this ambiguity originates from the six-fold symmetry of early diffraction patterns and the consequent classification of such a Ga2O3 phase as “hexagonal”, hence the ε-Ga2O3 ascription. A comprehensive TEM study,20 as well as other works recently published,21–24 showed definitely that the so-called κ phase is indeed orthorhombic and not hexagonal. In analogy to the usual nomenclature of other metal-oxides, it is therefore correct to denominate the orthorhombic phase as κ, as it will be done in this paper. It was also demonstrated that κ-Ga2O3 exhibits a spontaneous polarization and a ferroelectric character,25 that may turn out to be advantageous for the formation of a 2D electron gas in Ga2O3-based heterostructures.26,27
In the field of Ga2O3-based X-ray detectors, only amorphous and nanocrystalline films or β-Ga2O3 bulk crystals (Fe-doped or undoped) have been tested so far,28–32 but no investigations have been carried out on other Ga2O3 polymorphs. This, along with the very high resistivity measured on undoped κ-Ga2O3 grown by MOVPE, and the easy and cost-effective film deposition on a wide range of large-area commercial substrates (e.g., sapphire, GaN, and SiC), motivated the present investigation on the suitability of κ-Ga2O3 for X-ray detection. It is worth stressing here that although metastable, the κ phase is stable up to about 700 °C (then it fully converts to β at 900 °C33–35), which is sufficient for the standard fabrication of ohmic and rectifying contacts, as well as for employing κ-Ga2O3-based devices under standard operating conditions.
In this work, we report for the first time on the photoresponse of κ-Ga2O3 epitaxial thin films to X-rays. Along with a very low dark current, the results point out a high detection sensitivity, linearity with dose rate, fast response, and stability with time even at very low applied electric fields. It will be shown that some of these properties are superior to those previously reported for X-ray detectors based on amorphous,28 nanocrystalline,29 and bulk Fe-doped β-Ga2O3 photoconductors,30,31 or bulk undoped β-Ga2O3 Schottky diodes.32 These results are very encouraging, and pave the way to the application of orthorhombic undoped κ-Ga2O3 epitaxial thin films to direct X-ray detection.
Active material | Thickness (μm) | Bias voltage (V) | Operating electric field (V μm−1) | Rise time (s) | Responsivity (μC Gy−1) | Normalized sensitivity (μC mGy−1 cm−3) | Reported stability (min) | Year | Ref. |
---|---|---|---|---|---|---|---|---|---|
a-Ga2O3 | 0.25 | 5–100 | 1–20 | 15.5 | — | 271 | 20 | 2019 | 24 |
nc-β-Ga2O3 | 0.2 | 10–200 | 0.05–1 | 0.035 | 0.68 | 138.8 | 15 | 2021 | 25 |
sc-β-Ga2O3 (Fe) | 1000 | −1000 to −200 | −1 to 0.2 | <0.2 | 6.75 | 0.75 | 1 | 2021 | 27 |
sc-β-Ga2O3 (undoped) | 1000 | −25 to 0 | 0.025–0 | 13.8 | 1.8 | — | 5 | 2019 | 28 |
κ-Ga2O3 | 0.44 | 1–500 | 0.001–0.5 | <0.5 | 2.66 | 342.3 | 120 | 2022 | This work |
Prior to the deposition of the metal contacts, aimed at removing particulates and organic contaminants, the sample was cleaned in an ultrasonic bath in hot acetone for 5 min, then dipped in isopropanol for further 5 min, rinsed in deionized water, and finally dried in a pure N2 flow. Then, metal contacts were fabricated in a single step by RF magnetron sputtering deposition of a 300 nm thick Au film. The contact geometry (two rectangular 4.2 × 1.6 mm2 pads separated by a 1.0 mm-wide gap) was directly defined during deposition by the use of a stainless steel shadow mask positioned on the sample surface, with no need for photolithographic steps. The background pressure was 10−6 mbar, the target-to-substrate distance was 70 mm, and Ar was used as a sputter gas at a pressure of 2.5 × 10−2 mbar. Finally, the detector was mounted on a standard printed circuit board (PCB) equipped with a SMA connector for biasing and current signal collection (Fig. 1(b)).
Spectral UV-Vis-NIR photoconductivity analysis in the 200–1000 nm range, used to investigate the role of defect states in photoconduction mechanisms, was performed by measuring the photocurrent generated by a monochromatic radiation focused on the 1 mm-wide gap between the two Au contact pads. Light came out from the output slit of a Newport Cornerstone 260 monochromator coupled to a Hg–Xe lamp. Aimed at reaching a higher sensitivity under sub-bandgap radiation, modulated measurements were performed. For this purpose, light was chopped at a frequency of 14 Hz by a SR540 chopper, and photogenerated current signals were converted using a low-noise transimpedance amplifier (Princeton Applied Research 181) into a voltage measured using a lock-in amplifier (EG&G 5209). Bias voltage was applied by means of a Keithley 487 electrometer, used as a voltage source only.
Fig. 2 reports the measured dark current Id values as a function of the bias voltage VBIAS, ranging from 0.1 to 500 V (corresponding approximately, by considering a 1 mm-wide interelectrode gap, to an applied electric field F varying in the range of 10−4–0.5 V μm−1). As can be seen from Fig. 2(a) (linear plot), Au contacts show excellent linearity in the full range of applied bias voltage, with the absence of undesirable non-linear effects triggered at high VBIAS values. Moreover, as can be inferred from Fig. 2(b) (log–log plot), linearity with the bias voltage extends for more than three decades, pointing out the ohmic behaviour of Au metal contacts on undoped κ-Ga2O3, differently from what was recently reported on n-type ε-Ga2O3, on which a Schottky junction was formed.39 We have to indeed consider the very high resistivity of undoped κ-Ga2O3: this implies that Fermi energy is located deep into the bandgap, and that the material is fully depleted in practice, leading to negligible band-bending at the metal–semiconductor interface, i.e., to the formation of an ohmic contact. It is also worth mentioning here that both ohmic and Schottky contacts demonstrated to be effective for κ-Ga2O3-based MSM detectors:40 in particular, ohmic contacts should be preferred to enhance the device responsivity, whereas Schottky contacts are better when the response speed is a key concern. In our case, Au ohmic contacts ensure extremely low dark current values, ranging from a few fA for VBIAS < 1 V (Fig. 2(b)) to a maximum value of 6.7 pA at VBIAS = 500 V (Fig. 2(a)), which confirms the very high resistivity ρd of the deposited epitaxial κ-Ga2O3 thin film. From the evaluation of the slope of the linear best-fit shown in Fig. 2(b), it is possible to obtain ρd = 3.15 × 1010 Ω cm, which exceeds by more than one order of magnitude the threshold requirement for semiconductor-based X-ray detectors (108–109 Ω cm).41
Fig. 3 shows the time-dependent net photocurrents measured under X-ray on/off cycles (duty cycle 50%, time period 240 s) at low (1–20 V, Fig. 3(a)) and high (20–500 V, Fig. 3(b)) bias voltages, by progressively decreasing the dose rate at each “X-ray on” period. The net photocurrent Iph is given by Iph = Itot − Id, where Itot is the total current measured under irradiation. As expected, we can observe that: (1) at a given applied bias voltage, Iph increases with increasing dose rate; (2) at a given dose rate, Iph increases with increasing bias voltage. It is particularly worth noting the presence of a detectable signal even at VBIAS = 1 V, which corresponds to an applied electric field of only 10−3 V μm−1. Interestingly, the shape of the pulses is considerably different in the two operating regimes, as can be better appreciated in Fig. 4(a), showing a zoomed image of all the X-ray photocurrent pulses recorded in the bias voltage range of 1–500 V at the exemplary dose rate DR = 164 μGy s−1. It can be seen that at low bias voltages (1–20 V) the photocurrent fairly retraces the X-ray pulse excitation: the signal rapidly rises when the X-ray tube is switched on, then immediately settles down to its stationary value, and finally falls quickly to zero when the tube is switched off. Conversely, at high bias voltages (VBIAS > 20 V), after an initial overshoot followed by a rather fast decay, the signal increases very slowly towards its stationary value, which, however, is not even reached at the end of the “X-ray on” period (120 s); finally, an exponentially decaying residual signal is present after the tube is switched off.
By considering the stationary X-ray net photocurrent values (extrapolated in the case VBIAS > 20 V) obtained from Fig. 3 at the different applied bias voltages, it is possible to analyze the behavior of the induced signal with the delivered dose rate (Fig. 4(b)). As can be seen, the device exhibits a fairly linear photoresponse with dose rate. The dependence perfectly follows the Fowler's model42 for electrical conductivity of solid-state detectors, according to which Iph ∝ DRΔ, where Δ is the linearity coefficient. Values of Δ around unity are always desirable, even mandatory in the case of applications where linearity with the delivered dose rate is essential, such as dosimetry for radiotherapy. In the case of the device investigated in this work, Δ increases with increasing bias voltage in the range Δ = 0.76–0.99 (see inset of Fig. 4(b)). It is known from the literature on ionizing radiation detectors based on wide-bandgap semiconductors43 that the value of Δ is strongly influenced by the presence of defects, the role of which in the dynamics of X-ray photogenerated charge transport is in turn significantly dependent on the intensity of the applied electric field. Indeed, when defects cause trapping of the photogenerated carriers, charge collection efficiency unavoidably decreases, unless the applied electric field is intense enough to induce detrapping, thus compensating for the temporary signal loss. This is most probably what happens in the case of κ-Ga2O3. At low fields, charge trapping results in a slightly sub-linear photoresponse with dose rate (Δ = 0.76–0.88 in the 1–10 V range), but linearity progressively improves at higher fields (Δ = 0.94–0.99 in the 20–500 V range), approaching the ideal value Δ = 1.
The fairly linear relationship between Iph and DR allowed us to estimate the surface specific sensitivity of the device, defined as S = Iph/(A·DR), where A = 4.2 × 4.2 mm2 is the active area of the detector. The estimation was performed by fitting the photocurrent data reported in Fig. 4(b) with linear curves (not shown), and then evaluating the different slopes as a function of the bias voltage. Fig. 5 reports the values obtained for S, ranging from 0.50 ± 0.04 nC mGy−1 cm−2 at VBIAS = 1 V to 15.06 ± 0.27 nC mGy−1 cm−2 at VBIAS = 500 V. Note that the relative uncertainty decreases with increasing bias voltage, owing to the better linearity of the photoresponse. If we consider the film thickness (440 nm), the normalized sensitivity to the active volume of the detector ranges from 11.42 ± 1.02 μC mGy−1 cm−3 at VBIAS = 1 V to 342.29 ± 6.24 μC mGy−1 cm−3 at VBIAS = 500 V; it is worth stressing here that the latter is an unprecedented value for Ga2O3-based X-ray detectors.28–32 Responsivity, defined as R = Iph/DR, ranges from 0.09 ± 0.01 μC Gy−1 at VBIAS = 1 V to 2.66 ± 0.05 μC Gy−1 at VBIAS = 500 V, and is comparable to values reported in the literature for bulk single-crystal β-Ga2O3.31,32
Fig. 5 Surface specific sensitivity of the fabricated X-ray κ-Ga2O3 detector as a function of the bias voltage. Error bars are related to the linear fitting of the experimental data shown in Fig. 4(b). |
Unfortunately, the price to pay for a better linearity with dose rate and specific sensitivity at higher electric fields (in addition to the application of a high bias voltage, which is not desirable in any case) is a significantly longer time for the detector to reach a steady-state condition under irradiation, as it clearly appears from the 200 V and 500 V curves of Fig. 3(b) and 4(a). As can be seen, at high VBIAS, a photocurrent overshoot is triggered immediately after the irradiation starts, the amplitude of which increases with both increasing dose rate and bias voltage, i.e., both when a large amount of charge carriers is photogenerated (at higher DR) and is effectively collected by the electrodes (at higher VBIAS).
Photocurrent overshoots at the beginning of X-ray irradiation have already been observed in β-Ga2O3-based detectors,29 but no possible explanation was provided. However, transient photocurrent peaks appearing in a wide range of photoconductive devices, from diamond-based X-ray detectors44 to organic solar cells45 and light-emitting diodes,46 have been explained in the literature by trap-mediated space-charge effects. This is most likely what happens also in the case of κ-Ga2O3 at high bias voltages. At the beginning of the “X-ray on” period, the very fast rise of the photocurrent is due to the large amount of collected photogenerated carriers; however, we suppose that part of them gets trapped beneath the electrodes, causing space charge formation. As a consequence, an internal polarization electric field builds-up opposing the external bias, so that the charge collection efficiency is reduced, making the photocurrent decrease after peaking. A steady-state photocurrent should be reached at the end of the overshoot, soon after an equilibrium is established when polarization settles, but in our case the overshoot is followed by a very slow increase of the photocurrent, eventually exceeding (at the end of the “X-ray on” period) the overshoot peak value. This behavior may be explained by supposing that charge detrapping is triggered in the high-bias regime. To support this hypothesis, it is worth noting that the increase of photocurrent after the overshoot is more pronounced at higher bias voltages and higher dose rates, i.e., when the electric field is strong enough to promote charge detrapping and at the same time the radiation intensity is high enough to induce a significant trap-refilling.
A quantitative analysis of the timescales of the peak decay and the slow increase of photocurrent after the overshoot was performed, aimed at shedding light on the involved mechanisms. Two values of bias voltages were considered, namely, 200 V and 500 V. Curve fitting to the experimental data was performed by using the sum of two exponential functions, in the following form:
(1) |
Fig. 6 Curve fitting to experimental data of X-ray photocurrent at VBIAS = 500 V and (inset) VBIAS = 200 V. Dashed lines represent the best fit curves obtained by using eqn (1). For the sake of simplicity, the overshoot peaking time was assumed to be t = 0. |
The obtained values, averaged over all the dose rates in the range of 22.3–189 μGy s−1 (where the overshoot is present), are τ1 = 2.13 ± 0.07 s and τ2 = 32.48 ± 0.57 s for 200 V, and τ1 = 3.57 ± 0.09 s and τ2 = 31.82 ± 0.41 s for 500 V. Such values are consistent with the time constants of charge trapping/detrapping mechanisms reported for both β-Ga2O3- and κ-Ga2O3-based MSM photodetectors for UV radiation.47,48 It is worth recalling here that trapping/detrapping time constants are governed by the energy depth of the traps,48 and that longer time constants are related to deeper traps. As can be inferred from our results, the timescale (few seconds) of the peak decay (τ1) is consistent with charge trapping by shallow defects, responsible for the space charge formation after peaking, whereas the very high values (few tens of seconds) of τ2 justify a predominant detrapping from deep levels in the second stage of the transient regime. Once trapping and detrapping rates are equal, and a constant electric field is established between the collecting electrodes as well, a steady-state photocurrent is eventually reached.
To further support our hypotheses on the role of sub-bandgap defects on charge transport, spectral photoconductivity measurements were performed in the 200–1000 nm range. Fig. 7 reports the spectral responsivity SR of the κ-Ga2O3 epitaxial thin film measured at the same bias voltages (1–500 V) used for the X-ray characterization session; SR was calculated as a function of the impinging wavelength λ as SR(λ) = |Iph(λ)|/P(λ), where |Iph| is the modulus of the measured modulated photocurrent, and P is the power of the monochromatic radiation focused on the active area of the device. As expected, responsivity increases with increasing bias voltage, denoting a progressively better charge collection efficiency. In addition, all spectra show the presence of shallow traps, appearing as band-tails for λgap < λ < 320 nm, where λgap ≈ 270 nm is the wavelength corresponding to the energy bandgap (≈4.6 eV). More interestingly, note that photoconduction mediated by deep traps (λ > 320 nm) is boosted in the high-bias regime, increasing considerably at VBIAS = 50, 200, and 500 V; this experimentally confirms our hypothesis on detrapping from deep levels as the predominant photoconduction mechanism acting in the second stage of the X-ray photocurrent transient regime.
In addition to sensitivity and linearity with the dose rate, another valuable figure of merit for X-ray detectors is the response speed. By taking a closer look at the shape of a photocurrent pulse recorded during a single “X-ray on” period (Fig. 8), it is possible to evaluate the response speed of the κ-Ga2O3 film in terms of rise time tr (defined as the time required for the pulse to rise from 10% to 90% of its peak value) and decay time td (defined as the time required for the pulse to decay from 90% to 10% of its peak value). At low bias voltages (1–10 V), tr and td are approximately equal, and vary in the range of 1–2 s; in particular, both tr and td decrease slightly with increasing applied electric field (see Fig. 8(a) and (b)). Both the measured values (in the range of a few seconds) and the dependence on the bias voltage are in agreement with the typical electron/hole trapping effect, already reported in the literature for β-Ga2O3.49 A higher bias voltage leads indeed to a faster charge collection and reduces the charge trapping probability, decreasing both the rise and the decay time; in this sense, narrowing the electrode spacing (by using interdigitated contacts, for example) may be surely helpful for speeding up the photoresponse in the low-bias regime. When VBIAS = 20 V, tr and td are both lower than 0.5 s; however, it is worth stressing here that this value should be considered as an upper limit for our device, because it represents the minimum delay time between two consecutive measurement points obtainable with the electrometer used for the experiments. Indeed, as recently reported on X-ray detectors based on ε-Ga2O3 epitaxial thin films grown by MOCVD,50,51 and tested with a faster front-end electronics, the actual rise and decay times may be significantly lower (<50 ms).
As can be inferred from Fig. 8, the biasing condition VBIAS = 20 V ensures the best performance in terms of response speed and recovery time; the signal reaches its stationary value soon after X-ray irradiation starts, then immediately falls to zero when X-ray is shut down. At higher bias voltages (VBIAS > 20 V), pulse rising and decaying dynamics gets progressively slower (see Fig. 8(c) and (d)). In this operating regime, the applied bias is indeed high enough to trigger both space charge formation and detrapping, thus requiring longer times to obtain a stable photocurrent when X-rays are on, as well as to recover the initial conditions at the end of the irradiation period. We can therefore deduce that VBIAS = 20 V represents an optimal operating condition for the fabricated device, ensuring at the same time a fast response (tr, td < 0.5 s) and a very good specific sensitivity (S = 5 nC mGy−1 cm−2) to X-rays; most probably, the corresponding applied electric field is in this case high enough to ensure a fast and efficient charge collection, and at the same time low enough to avoid both photocurrent “crowding” beneath the electrodes (thus hampering space charge formation) and detrapping from deep levels.
Operation of the device at the optimal bias voltage (20 V) was then used to verify the signal reproducibility at a given dose rate, as well as to preliminarily assess the detector's response stability with time. Fig. 9 reports the X-ray photocurrent signal recorded at VBIAS = 20 V in a 2 h measurement session at the highest investigated dose rate (189 μGy s−1). In the first part of the session, the device was tested with 10 X-ray on–off cycles (duty cycle 50%, time period 360 s), and the recorded signal pulses demonstrated an excellent reproducibility. Then, the device was kept for 60 min under the maximum X-ray flux, producing a perfectly stable signal, only varying within ±1.5% of its average stationary value (165.5 pA, see Fig. 9).
X-ray irradiation was finally profitably used to evaluate the average mobility-lifetime product μe,hτe,h of the photogenerated carriers, which is the most significant figure of merit for the charge transport properties of a semiconductor, being directly related to the concentration of defects acting as recombination centers for both electrons and holes. (i.e., it provides an estimate of the crystalline quality of the deposited film). A simple way to extract the μe,hτe,h product is by fitting the X-ray photocurrent Iphvs. VBIAS curve according to Hecht's equation:52
(2) |
Fig. 10(a) reports the experimental results obtained by irradiating the fabricated detector at different dose rates. Best fits returned μe,hτe,h = (1.91 ± 0.07) × 10−4 cm2 V−1 at DR = 189 μGy s−1, μe,hτe,h = (2.37 ± 0.08) × 10−4 cm2 V−1 at DR = 91.9 μGy s−1, and μe,hτe,h = (3.13 ± 0.12) × 10−4 cm2 V−1 at DR = 45.3 μGy s−1. The first result to be noted is that all the estimated μe,hτe,h values are about one order of magnitude higher than those recently reported for single-crystal Fe-doped β-Ga2O3-based detectors tested under the same experimental conditions (type of X-ray source, accelerating voltage, and tube current).26 This is most probably due to a higher lifetime of the photogenerated carriers before recombination; indeed, high Fe-doping of β-Ga2O3, necessary to increase the material resistivity and lower the dark current, greatly reduces both the carrier drift length and the carrier collection efficiency, due to the introduction of deep-level impurities.53 Moreover, it is worth noting from Fig. 10(a) that μe,hτe,h decreases slightly with an increasing dose rate, presumably due to a higher density of photogenerated charge, enhancing carrier–carrier scattering effects which can limit carrier mobility.
Fig. 10 (a) Experimental results for the evaluation of the average mobility-lifetime product at different dose rates. Bias voltage was varied in the 1–500 V range. In all the cases, dashed lines show the best fit to data obtained by using Hecht's eqn (2); (b) the same plot as Fig. 10(a) zoomed in the 1–10 V range, aimed at showing the ohmic behaviour of the metal contact under X-ray irradiation. Dashed lines represent the best linear fit to experimental data. |
It is also interesting to observe that the ohmicity of the metal contact remains even under X-ray irradiation, as can be inferred from Fig. 10(b), showing a zoomed image (in the 1–10 V range) of the plot reported in Fig. 10(a). In this regime of applied bias voltages, in which the photocurrent saturation is not started yet, the exponential term of Hecht's equation can be indeed neglected to a first approximation, and the relationship between Iph and VBIAS is linearized. In this way, the contact ohmicity can be correctly verified, as in our case. Moreover, it can be inferred from Fig. 10(b) that the slope of the linear fitting curve increases with increasing dose rate, as expected from a photoconductive device, the conductivity of which is directly proportional to the intensity of the absorbed radiation.
It is worth highlighting that κ-Ga2O3 shows the highest normalized sensitivity to the active volume, and is the only polymorph which demonstrates excellent stability under 120 min of X-ray irradiation, without showing any degradation of the photoresponse. It shares with amorphous (a-) Ga2O3 and nanocrystalline (nc-) β-Ga2O3 the advantage, if compared to β-Ga2O3 single crystals, of being deposited as a thin film, allowing for the fabrication of large-area devices. It is slightly slower than nc-β-Ga2O3, but significantly faster than a-Ga2O3, and more responsive than both in a wider operating voltage range. Fe-doped single-crystal (sc-) β-Ga2O3 shows better properties than κ-Ga2O3 in terms of rise time and responsivity, but at the expense of very high bias voltage, which makes it unsuitable, for instance, for the implementation of low-power portable devices; conversely, undoped sc-β-Ga2O3 can operate at low bias voltages, but it shows both a slower response and a lower responsivity than κ-Ga2O3.
In conclusion, in comparison with the other members of the gallium oxide family, κ-Ga2O3 demonstrated to be the most versatile candidate for the fabrication of direct X-ray detectors, the performance of which can be tailored to the specific user's requirements, depending on the operating voltage. High bias regimes (VBIAS > 20 V) are to be preferred to maximize sensitivity and linearity with X-ray dose rate regardless of the response speed, but low bias operation (down to 1 V) is also possible, with the advantage of a very low power consumption. In this work, it is shown that an intermediate bias voltage (VBIAS = 20 V) is the optimal solution, ensuring the best trade-off between sensitivity and linearity with the dose rate, as well as a fast and rapidly stable photoresponse. Anyway, it is finally worth mentioning that, by reducing the gap between the electrodes at values lower than 1 mm (e.g., by the use of interdigitated electrodes fabricated by photolithographic techniques), it is possible to further reduce the value of the applied bias voltage while maintaining the same charge collection performance.