Gerard
Macias‡
a,
Justin R.
Sperling‡
a,
William J.
Peveler
b,
Glenn A.
Burley
c,
Steven L.
Neale
a and
Alasdair W.
Clark
*a
aSchool of Engineering, University of Glasgow, Rankine Building, Oakfield Avenue, Glasgow, UK. E-mail: Alasdair.clark@glasgow.ac.uk
bSchool of Chemistry, University of Glasgow, Joseph Black Building, Glasgow, UK G12 8QQ
cWestCHEM & Department of Pure & Applied Chemistry, University of Strathclyde, 295 Cathedral Street, Glasgow, G1 1XL, UK. E-mail: glenn.burley@strath.ac.uk
First published on 3rd July 2019
Metallic nanostructures are ideal candidates for optical tongue devices thanks to their chemical stability, the sensitivity of their plasmonic resonance to environmental changes, and their ease of chemical-functionalization. Here, we describe a reusable optical tongue comprising multiplexed gold and aluminum nano-arrays: a bimetallic device which produces two distinct resonance peaks for each sensing region. Through specific modification of these plasmonic arrays with orthogonal surface chemistries, we demonstrate that a dual-resonance device allows us to halve sensor sizes and data-acquisition times when compared to single-resonance, monometallic devices. We applied our bimetallic tongue to differentiate off-the-shelf whiskies with >99.7% accuracy by means of linear discriminant analysis (LDA). This advance in device miniaturization, functionalization, and multiplexed readout indicates nanoplasmonic tongues will have future applications in chemical mixture identification in applications where portability, reusability, and measurement speed are key.
Chromatography is the current gold-standard for detection, identification, and classification of chemical components from complex gas12 and liquid13 mixtures. However, the nature of chromatographic identification techniques (such as liquid chromatography mass spectrometry) requires specialized laboratory equipment for the separation and analysis of a sample's chemical components. This results in costly, time-consuming, and often low throughput processes14 that are unsuitable for applications where real-time monitoring or portability are desirable (air and liquid sampling in the security, food, or drug sectors, for example). In response to these issues artificial ‘tongues’ and ‘noses’ consisting of multiple, cross-reacting sensing elements have been developed.10,15–19 Compared to the aforementioned laboratory-based tools,12,13 these devices are portable, highly sensitive,20 do not require component isolation, and can be fabricated relatively cheaply.21,22
Our perceptions of taste and smell rely on multiple partially-selective chemoreceptors that result in distinct electro-chemical patterns for specific flavors and odors.23 Influenced by this mechanism, artificial tongues/noses work by combining the responses of multiple cross-reactive sensors, allowing them to identify chemical mixtures through trained pattern recognition15,20 rather than by measuring absolute concentrations of specific components within the mixture.20,24,25 The more sensing regions added to the analysis, the more potential the device has to differentiate between mixtures.26 Pattern-recognition sensors such as these are extremely versatile and have found applications in medical diagnostics,16,27–31 environmental monitoring,32,33 and food-safety.10,17,34
A wide variety of materials have been used in order to construct artificial pattern-recognition sensors, including fluorescent polymers, doped metal oxides, and olfactory proteins.25,35,36 Recently, surface formed plasmonic gold (Au) nanostructures have emerged as a particularly useful sensing platform for these systems thanks to their chemical stability, the sensitivity of their plasmonic resonance to environmental changes, their ease of chemical-functionalization, and their reusability in comparison to solution phase sensing arrays.37,38 The optical response of Au nanostructures is dictated by their localized surface plasmon resonance (LSPR), a phenomena particularly sensitive to changes in local refractive index.39 Partial-selectivity in these devices is achieved by using multiple arrays of nanostructures, each modified with a different surface-chemistry, as individual sensing regions. When exposed to the same solution, the resonance peak-shift of each region varies due to the particular local chemical modification. Monitoring these variations results in the desired “fingerprint” for that mixture.31 However, the need for multiple sensing regions inevitably impacts device size and measurement times. As a result, there are size, weight, and speed advantages associated with reducing the number of sensing regions required for mixture classification.
Here, we present a reusable optical tongue device with three sensing regions, where each region is capable of obtaining two partially selective responses from a single measurement. Each region consists of two superimposed nanoplasmonic arrays featuring two distinct metals: Au and aluminum (Al). This allows for the orthogonal chemical-functionalization of each superimposed array via thiol (Au)40–42 and silane (Al)41,43 chemistries, while also allowing us to obtain two resonance peak-shifts using a single optical measurement. Compared to a device containing its monometallic counterparts, we demonstrate that our device containing bimetallic Au/Al sensors can halve the number of sensing elements required (reducing device size and number of regions to probe [i.e. data acquisition time]) without compromising the identification and classification capabilities of the device. We go on to show that these sensors can be used as an optical tongue to distinguish between seven different whiskies and three controls.
Both Au and Al can support selective functionalization of their surfaces. While the Au nanostructures can be readily modified by thiol chemistry,40–42 the native oxide layer present on the Al nanostructures displays –OH groups which enables modification via silane chemistry.41,43 The presence of organic ligands on plasmonic arrays is known to influence the extent to which certain organic molecules interact with the arrays, thus affecting the refractive index around the nanostructures and in-turn their resonant properties.37 While monometallic sensor arrays with single-ligand modifications have been reported,31 bimetallic arrays that allow dual-ligand modifications have yet to be explored.
Our system comprised of 3 bimetallic sensor arrays, each exhibiting unique surface chemistries: a sensor consisting of native Au and Al (Fig. 2a(i)); a sensor where the Au was modified with DT (1-decanethiol) and the Al with HMDS (hexamethyldisilazane) (Fig. 2a(ii)); and a sensor where the Au was modified with PFDT (1H,1H,2H,2H-perfluoro-1-decanethiol) and the Al with PEG (2-[methoxy(polyethyleneoxy)6–9propyl] trimethoxysilane) (Fig. 2a(iii)). These surface chemistries were chosen to represent varied levels of hydrophobicity/philicity and different chemical functionalities. Altering the surface chemistry of the nanostructures affects how individual chemical components in a mixture interact with the structures, altering their optical response. In addition to this “bimetallic” sensor array, a corresponding array of 6 equivalent monometallic sensors of Au and Al were also produced, matching the chemistries used on the bimetallic sensors (i.e. 3 Au arrays and 3 Al arrays). The monometallic (Fig. 2b) and bimetallic (Fig. 2c) sensors were tested against varying refractive index media adjusted with (i) acetone and (ii) ethanol. The resulting resonance shifts from water (RIU = 1.333) were compared using RIU values for acetone1 and ethanol8 solutions.
Fig. 2 Effect of surface chemistry on the sensitivity of Au, Al, and Au/Al sensor arrays. (a) Surface chemistry combinations used: (i) native Al, Au (ii) Al-HMDS, Au-DT, and (iii) Al-PEG, Au-PFDT. (b) The shift in plasmonic response from water for monometallic arrays in 10%, 20%, and 30% solutions (v/v) of (i) acetone and (ii) ethanol. (c) The shift in plasmonic response from water for bimetallic arrays in 10%, 20%, and 30% solutions (v/v) of (i) acetone and (ii) ethanol. The different surface chemistries (native Al, Al-HMDS, Al-PEG, native Au, Au-DT, and Au-PFDT) alter the plasmonic peak of the nanostructures when exposed to the same organic solvent. This results in different peak-shifted curves. The RIU values for acetone and ethanol solutions were obtained from S. S. Kurtz, et al. (1965)1 and T. A. Scott (1946),8 respectively. For (b) and (c), the lines are present to guide the eye and the error bars are one standard deviation from the average. |
Three trends were identified:
(1) Regardless of the metallic composition of the nanostructures, the organic solvent used to modify the refractive index, or whether the region is monometallic or bimetallic, the sensitivity curve depends on the organic ligand present on the nanostructure (e.g., the Al, Al-HMDS, and Al-PEG curves in Fig. 2b(i) are different).
(2) For any given surface chemistry on either the monometallic or bimetallic sensor, the sensitivity curve depends on the type of organic solvent used to alter the refractive index rather than just shifting with RIU alone (e.g. the Al-HMDS curves in Fig. 2b(i) and (ii) are not the same.)
(3) The sensitivity curves of the monometallic and bimetallic sensor for the same metal composition, organic ligand, and organic solvent, differ; the bimetallic sensors response is fundamentally different from its monometallic counterparts (e.g. the Al-HMDS sensitivity curves in Fig. 2b(i) and c(i) are not the same.)
In all three cases, we attribute these behaviors to the segregation of the solvent at the sensor-liquid interface and corresponding changes to the local refractive index. Solvent segregation is dependent on the chemical groups present at the interface;45 using different metals and different ligands on the surface results in different segregation behavior, which likely explains the different plasmonic responses. This is especially important when comparing the monometallic and bimetallic responses; the presence of a second metal and second ligand results in additional differential solvent segregation behavior. These results confirm that we can tailor the partial selectivity of the device via the orthogonal silane and thiol chemistry.
To further verify the applicability of the bimetallic approach for implementation as an optical tongue device, we performed a principal component analysis (PCA – a non-biased, multivariant analysis technique)19,46 across 10 different bimetallic ‘tongues’ using the data from our acetone/ethanol test. Each bimetallic tongue consisted of three element pairs: Al/Au, Al-HMDS/Au-DT, and Al-PEG/Au-PFDT. For the PCA, each row corresponded to a particular solvent tested, and each column corresponded to the transmission peak minimum for each surface chemistry. Further explanation of the PCA analysis used can be found in the Data Matrix for Analysis section of the ESI.†
Fig. 3 shows the PCA of the first two principal components (that explain 87.3% of the total variance), where black dots represent DI water, red dots represent acetone-based media and yellow dots represent ethanol-based media. These results show that by combining the response of multiple surface chemistries on our bimetallic sensors, we are able to cluster the results from each solution into a map.
While delineation of classes (acetone/ethanol and the v/v percent of each) is shown, it is important to note that this PCA analyzed the results across 10 different optical tongue devices. A close look at SEM images of each of these devices revealed that, while within the specifications of our e-beam lithography tool (i.e. 20 nm spatial resolution), the X–Y distances between our two metals was slightly different in each device (see ESI†). Given the high sensitivity of plasmonic nanostructures to their near-field environment, such minute misalignments can result in slight differences from sensor to sensor47,48 (confirmed by simulations Fig. S3, ESI†). Additionally, variations in position between the Al and Au arrays can alter the surface wettability and segregation properties. This is because the distribution of hydrophobic and hydrophilic groups is dependent on the position of the metals and their uniquely modified surface chemistries within the array.49 Thus, the spread of points within each class in the PCA is most likely attributed to this fabrication resolution. Regardless, the PCA shows ordering of the different solutions by combining the response from three sensing regions, which constitutes the basic requirement for the development of an artificial tongue. Similar behavior was observed with comparable monometallic sensors (six sensing regions) and can be found in the ESI.†
To further demonstrate the capabilities of the bimetallic tongue, we used one device to differentiate between seven different whiskies with identical ethanol contents (40%), a 40% vodka, and 40% ethanol in water, with water as the control (as shown in Table 1). This test was performed on a single bimetallic array to minimize the variance between sensors that would increase the noise within the data. The resulting response of the bimetallic array was compared to an equivalent monometallic array (containing six sensing regions). Fig. 4a(ii) and b(ii) show the 2D PCA results for the monometallic and bimetallic tongues, respectively. These PCAs show only the alcoholic solutions; water is off the axis set. Full PCAs can be viewed in the ESI.†
Fig. 4 Monometallic versus bimetallic tongues for whisky differentiation. Results for (a) monometallic and (b) bimetallic sensor arrays for the differentiation of the solutions found in Table 1. For each full tongue device: (i) SEM showing the make-up of each tongue device, (ii) PCA zoomed in to show only the alcoholic solutions; (iii) the PCA scree plot; and (iv) LDA zoomed in to show only the alcoholic solutions. The confidence ellipses for the LDA are one standard deviation. The colors used in the PCA and LDA are identified by the first column of Table 1. Full PCA and LDA figures can be found in the ESI.† |
Sensor performance is determined by the dimensionality of the PCA, the distance between the groupings, and ‘tightness’ of the groupings. The dimensionality is measured by the number of components required to account for 95% of measurement variance, as shown in Fig. 4a(iii) and b(iii). For the plasmonic tongue comprised of six monometallic sensors, two dimensions (principal components) contained >95% variance; and for the plasmonic tongue comprised of three bimetallic sensors, the first two principal components (PCs) contained 94.6% variance (with >95% of the variance over three dimensions). The overall difference between the cumulative variance of monometallic and bimetallic tongues with two principal components is very small. In both cases, the important qualitative point is that the PCA algorithm shows distinct differentiation of the different test solutions with large spacings between these groupings.
In both PCA analyses (mono- versus bimetallic) the pattern of water versus whisky and ethanol/vodka versus whisky is largely similar. W1 (Glenfiddich® 12y) in particular gives a markedly different signal to the other spirits tested. Analysis of the PCs in each tongue give an indication of the elements contributing to the sensor response. As shown in Table S3 (ESI†), for the monometallic tongue, PC1 is from the transmission peaks corresponding to the Au nanostructures, particularly Au and Au-PFDT that separate water from ethanolic solutions. Al-HDMS contributes to the PC2, along with Al which has the most separation of the whiskies/controls. In the bimetallic tongue, many chemical functionalities contribute to the PC1, leading to the separation of the water and whiskies as well as improved separation between vodka and ethanol solution, but PC2 is dominated by Al and Al-PEG, demonstrating that by combining the surface chemistries in a single device, different behavior is observed.
In this sensor configuration we hypothesize that whilst the main driver in the solution fingerprints is clearly the EtOH content (vs. water), the trace differences between a pure EtOH solution and the compounds present in the spirits are causing significant signal variations. These compounds include the additional alcohols present in whiskies (propanols and butanols), organic aromatic components (phenols, terpenes and vanillin), and aliphatics (lactones). Each of these components will have different interactions with the sensor surface coatings dependent on their partial solubility and hydrophobicitiy/phillicity. It is proposed the most hydrophilic components will interact favorable with the bare Al and PEG surfaces, whilst the most hydrophobic will prefer to interact with the Au and Au-DT surfaces. Factors such as pH or ionic strength may also contribute to the subtle changes seen on the sensor chips.
After analyzing the PCA and discrimination capabilities of both mono- and bimetallic tongues, we can conclude that both tongues are able to differentiate between the whiskies tested thanks to the functional groups present on their surface. To investigate whether formal classification was possible, linear discriminant analysis (LDA), a supervised technique, was applied to the data to generate new “scores” (in a similar methodology to PCA) to maximize separation between known clusters whilst minimizing variance within each cluster.50 Both the mono- (Fig. 4a(iv)) and bimetallic (Fig. 4b(iv)) tongues performed excellently and could classify (using leave-one-out cross validation to test accuracy) 100% and 99.7% of the data, respectively. Although the bimetallic tongue confused one instance of W3 for W5, this was arguably compensated for by its ability to provide two signals from one measurement and therefore make half the number of measurements required to collect the data, requiring less sample volume. Additionally, the greater cross-reactivity on these sensor elements increases the potential for tuning and improving the fingerprint responses by incorporating different pairings of surface chemistry.
Footnotes |
† Electronic supplementary information (ESI) available: Simulations to determine nanostructure periodicity and the effect of the distance between metallic arrays in the bimetallic regions; SEM images of each of the fabricated devices; transmission response showing the effect of different surface chemistries on sensing devices in water; sample of the data matrix used for PCA and LDA analyses; PCA figures for organic solvent differentiation; PCA figures and PC coefficients for whisky differentiation; LDA figures for whisky differentiation. See DOI: 10.1039/c9nr04583j |
‡ These authors contributed equally to this work. |
This journal is © The Royal Society of Chemistry 2019 |