Hiroyuki
Shibata
,
Yuki
Hiruta
and
Daniel
Citterio
*
Department of Applied Chemistry, Keio University, 3-14-1 Hiyoshi, Kohoku-ku, Yokohama 223-8522, Japan. E-mail: citterio@applc.keio.ac.jp; Fax: +81 45 566 1568; Tel: +81 45 566 1568
First published on 11th December 2018
Although the determination of calcium ions (Ca2+) is of high importance to monitor water hardness, currently available devices for on-site analysis suffer from a lack of user-friendliness and sensitivity. This work demonstrates fully inkjet-printed and low-cost microfluidic paper-based analytical devices (μPADs) for the simple naked-eye colorimetric determination of calcium ions (Ca2+) in drinking and tap water samples. The quantification of Ca2+ relies on visual readout of the length of a colour-changed detection channel modified with ionophore-doped ion-selective optode nanospheres (nano-optodes), eliminating the requirement of a scanner or a camera. All fabrication steps for deposition of assay reagents have been performed by means of a simple desktop thermal inkjet printer, which is expected to contribute to highly batch-to-batch reproducible device preparation. The detectable Ca2+ concentrations between 0.05 mmol L−1 and 5 mmol L−1 cover the range recommended by the International Organization for Standardization (0.05–2.5 mmol L−1) and the World Health Organization (WHO) guideline for Ca2+ quantification in drinking water (less than 5 mmol L−1). The lowest concentration of Ca2+ detectable by the naked eye was found to be 0.05 mmol L−1, which is below the value achieved with previously reported paper-based devices. μPAD quantified Ca2+ concentrations in tap or drinking waters were within 15% error of the results obtained with a classical complexometric titration. Hence, distance-based μPADs relying on nano-optodes are sensitive and reproducible tools for equipment-free on-site assaying of Ca2+ in real samples.
Since Whitesides’ research group first established the concept of microfluidically patterned paper,6,7 microfluidic paper-based analytical devices (μPADs) have emerged as a new class of analytical platform over conventional paper strips for point-of-need analysis.3,8–14 In this context, the naturally porous structure of cellulosic paper plays an important role as a sustainable, inexpensive, ready-to-use, and safely disposable substrate platform for (semi-)quantitative assays applied to both clinical and environmental monitoring. Passive sample transportation through capillary forces enables sophisticated (bio)chemical assays, including sample pretreatment steps, and μPADs have evolved into analytical tools helping to overcome some limitations of conventional analytical instruments in terms of cost, portability and user-friendliness. Colorimetric μPADs are of particularly high interest, because they allow for unaided eye observable quantitative signal readout.15 Not surprisingly, various metal ions (especially heavy metals) have been colorimetrically determined on μPADs due to the wide availability of classical colorimetric indicators.3,12–14,16–20 With regard to the colorimetric detection of Ca2+, complexometric titration reagents, such as chelating reagents, cation-responsive chromogens, and masking reagents, have typically been adapted to the cellulosic substrate.21–23 Kaneta's research group for example, fully integrated the classical complexometric titration system into a μPAD for the visual determination of Ca2+ and Mg2+ in natural or drinking waters without the requirement for burettes and optical signal quantification tools (e.g. camera, scanner).22 Nevertheless, paper-based assays for visual detection of Ca2+ in drinking water and tap water suffer from relatively poor limits of detection (LOD), with an LOD of 0.21 mmol L−1 reported for a μPAD relying on software-assisted colorimetric readout,21 or of 5 mmol L−1 by naked eye visual detection.22
As one of the sophisticated optical chemical sensors targeting ionic species, ionophore-based ion-selective optodes (ISOs) continue to attract significant attention.24–26 Classical ISOs are typically composed of lipophilic sensing components, such as a pH indicator as optical transducer (chromoionophore), an ion-specific ligand (ionophore), and an organic salt additive (ion-exchanger), embedded into a plasticised polymeric membrane.24,27 Due to the versatile capabilities of ISOs, their detection function has been integrated into cellulosic material platforms, such as a cotton thread28 and paper substrates.29–36 More recently, our research group introduced the classical polymer film-based ISOs into μPADs with the aid of inkjet printing technology.29 In a later stage, we reported the use of inkjet-generated polymeric particle-based ISOs on paper strips.30 Meyerhoff's research group on the other hand, first established a new class of ISOs without the conventional plasticised polymeric materials on paper substrates,31 and first applied inkjet printing technology for the deposition of the corresponding ISO components.32–34 Although paper-based ISOs have served to determine various electrolytes, they require optical signal readout tools (e.g. scanner, camera, reflectometry). Only in 2018, Henry's and Bakker's research groups succeeded in the elimination of the conventional colour intensity or colour hue readout by adapting a distance-based signal readout motif to ISOs with micelle-based nanospheres (nano-optodes) pipetted onto μPADs.35 Nevertheless, inkjet printing technology has to best of our knowledge never been applied to the highly-reproducible and well-defined deposition of all assay reagents onto practical μPADs.
The present work demonstrates distance-based μPADs for the naked-eye quantification of Ca2+ in drinking or tap waters with high batch-to-batch reproducibility and assay accuracy without any external equipment. For this purpose, nano-optodes composed of the surfactant Pluronic® F-127 have been applied to the fabrication of distance-based μPADs by means of inkjet printing technology. Notably, the nano-optodes no longer require the use of volatile organic solvents in the paper device fabrication process (e.g. tetrahydrofuran, cyclohexanone).25,37,38 Therefore, it is possible to use simple desktop thermal inkjet printing for their deposition onto wax-patterned paper substrates. All fabrication steps leading to the distance-based μPADs, from the microfluidic patterning of the paper substrate (wax printing) to the deposition of all required assay reagents (inkjet printing), were performed with standard printing technology, which is expected to contribute to highly reproducible and scalable fabrication with minimum consumption of chemical reagents.39,40 Additionally, the developed μPADs show a lower naked-eye observable LOD (0.05 mmol L−1) than a commercial colorimetric paper dipstick (0.1 mmol L−1), and they are applicable to practical sample analysis of Ca2+ with sufficient accuracy (no more than 15% error).
Ultrapure water (>18 MΩ cm) was obtained from a PURELAB flex water purification system (ELGA, Veolia Water, Marlow, U.K.) and used for the preparation of all solutions. Advantec No. 5C filter paper was purchased from Toyo Roshi (Tokyo, Japan). Hot lamination films (150 μm thickness, film material: polyethylene terephthalate and polyvinyl alcohol as a thermoplastic adhesive) were obtained from Jointex (Tokyo, Japan). For inkjet printing of the reagents for Ca2+ detection, a thermally-actuated Canon iP2700 inkjet printer (Canon, Tokyo, Japan) was used. For this purpose, the standard colour cartridge of the Canon printer was cut open and the sponge inside and colour inks were removed, followed by washing with copious amounts of deionised water. Dynamic light scattering (DLS) measurements to evaluate the hydrodynamic diameter of nano-optodes were performed by a Zetasizer Nano ZS from Malvern (Worcestershire, UK). Optical microscope images were acquired on a DVM2500 digital microscope (Leica, Wetzlar, Germany).
As sample pretreatment reagents, pH-buffering solution (250 mmol L−1 HEPES-TMAOH buffer, pH 7.0) and 10 mmol L−1 aqueous MgCl2 solution, were separately inkjet-deposited from a black ink cartridge in 1 and 3 print cycles, respectively. Then, Ca2+-selective nano-optodes were inkjet-printed from a magenta ink cartridge in 20 printing cycles. After the inkjet printing step, the printed sides of the μPADs were laminated with exception of the sample inlet area to limit evaporation of sample liquid during assays.
The appropriate amounts of chemical reagents for distance-based Ca2+ assays (e.g. pH-buffering reagents, pre-deposited MgCl2 salt) were investigated. First, the influence of the amount of deposited pH-buffering reagents (250 mmol L−1 HEPES-TMAOH buffer at pH 7.0) was evaluated. For this purpose, 4 different numbers of printing cycles (1, 2, 3 and 5 cycles) were tested with distance-based μPADs (results shown in Fig. S6†). The deposition of two or more printing cycles of the pH-buffer solution resulted in a high background signal indicated by the magenta colour observable also in the absence of Ca2+. This is probably because an excess quantity of cationic buffer components (i.e. TMA+) affects the cation-exchange reaction of the nano-optodes by an uptake of buffer cations into the nano-optode phase and hence, the deprotonation of chromoionophore even in the absence of target cations. Therefore, one single printing cycle was selected as the optimal amount of deposited pH-buffering solution in device fabrication. Next, the effect of the amount of deposited MgCl2 for improved sensitivity of μPADs was studied (refer to Fig. S7† for the quantitative response curves). In this work, MgCl2 was deposited in the inlet area to compete in the non-specific interaction between the target cations (Ca2+) and the cellulosic surface during capillary force-driven sample wicking, which results in improved sensitivity for the distance-based Ca2+ assay. The reason for this is further explained in the section discussing the interference study. MgCl2 has been selected because the used Ca2+ ionophore has excellent selectivity against Mg2+,41 so that the presence of this cation does not contribute to a colorimetric signal. The experimental results of Fig. S7† show that the response curves for Ca2+ are vertically shifted towards longer colour-changed distances with increasing amounts of printed MgCl2. However, excess amounts of deposited MgCl2 (i.e. over 5 printing cycles) resulted in a high background signal upon the corresponding μPADs being exposed to the blank and 1 mmol L−1 of Ca2+ samples (refer to Fig. S8†), due to the extraction of Mg2+ into the Ca2+-selective nano-optodes. Therefore, three printing cycles were selected as the appropriate amount of deposited MgCl2 in subsequent experiments.
Fig. 4 (a) Reproducibility evaluation of distance-based μPADs independently fabricated on different days for software-assisted Ca2+ quantification; each data point has been obtained by measurements with 4 individual single-use devices; 30 μL of aqueous CaCl2 solution applied; error bars indicate the standard deviations; incubation time: 45 min; circle plots (batch #1) are identical to the data of Fig. 3b; (b) response curves for distance-based Ca2+ detection obtained by scanner and software-assisted digital colour analysis (circles) and individual user naked-eye readout (crosses and triangles); each data point has been obtained by measurements with 3 individual single-use devices; 30 μL of aqueous CaCl2 solution applied; error bars indicate the standard deviations; incubation time: 45 min. |
To address the analytical capabilities of the μPADs, the lowest naked-eye detectable concentrations were compared with those achieved with commercial colorimetric paper dipsticks. The assay procedure applied for commercial paper dipsticks and the colorimetric changes corresponding to various Ca2+ concentrations are shown in Fig. S11.† In this study, the LOD was defined as the lowest concentration observable by naked eye as a visual colour change instead of the classical 3σ method, since a blank sample does not generate any optical signal on distance-based μPADs (i.e. blank signal distance = 0 mm). From Table S1,† it can be concluded that the developed distance-based μPADs resulted in lower LOD (0.05 mmol L−1) than commercial colorimetric paper dipsticks for the Ca2+ assay shown in Fig. S11† (0.1 mmol L−1). Furthermore, it should be noted that commercial paper dipsticks require a multi-step procedure, including sample pretreatment by reagent addition (Fig. S11a†). Thus, the distance-based μPADs represent sensitive and user-friendly tools for the instrument-free and single-step determination of Ca2+ concentrations.
Despite their improved sensitivity for selective naked-eye determination of Ca2+ in mineral or tap water, the proposed distance-based paper devices require a relatively long assay time (45 min) to acquire the quantitative optical signal. It is a general drawback that distance-based detection motifs need a prolonged time for assay completion due to the slow transport of sample liquid by capillary force: up to 45 min have also been reported by others for several centimetres of detection channel length.16,17,20 In addition, distance-based colorimetric signalling relies on analyte depletion during sample transport, either by binding to a specific receptor16 or by interaction with the paper substrate, as it is the case in the current work. This can only be achieved with sample liquids being slowly transported through the cellulose fibre network of the paper substrate.
On the other hand, the distance-based colorimetric signal achieved with the present system is stable over time (up to approximately 2 h) so that a user is not forced to continuously monitor the device during an assay, which allows for so-called “walk-away assays” that require nothing more than the application of the sample. While it is clear that in general a more rapid assay time is desirable, the 45 min assay time is not hindering practical use of the proposed devices. A possible perspective to reduce assay times is the shortening of flow channels. However, this would result in narrowing of the dynamic response range and reducing the signalling resolution. Alternatively, converting a colour intensity-based approach into a distance readout referred to as “dip-and-read distance-based assay” as we have recently reported,20 might allow nearly instantaneous signal detection, although probably at lower resolution. Furthermore, dipping time dependent result variations would have to be expected, due to the dissolution of buffer salts required for maintaining a constant pH value.
Interfering ions | Tolerance [mmol L−1] | Interfering ions | Tolerance [mmol L−1] |
---|---|---|---|
Cl− was used as counter anion for preparation of all samples. | |||
Na+ | 100 | Cu2+ | 1000 |
K+ | 100 | Ni2+ | 100 |
Li+ | 100 | Hg2+ | 100 |
Mg2+ | 100 | Al3+ | 1000 |
Zn2+ | 1 |
It is postulated that the adsorption of Ca2+ to the paper substrate is the primary cause for the analyte concentration-dependent changes in colour-changed distance, rather than the depletion of the target cation by extraction into the optode phase. To verify this hypothesis, various concentrations of pH-buffered Ca2+ samples were applied onto a reagent-free paper channel, and Ca2+-selective nano-optodes were inkjet-printed onto the μPADs only after exposure to Ca2+ samples (refer to the ESI and Fig. S14† for the detailed procedure and the quantitative response curve). The data of Fig. S14† clearly shows that a Ca2+ concentration dependent colour-changed length is achieved also in the absence of nano-optodes during sample flow through the detection channel. Hence, it can be concluded that distance-based quantification in this work relies on the non-specific adsorption of Ca2+ onto the paper substrate. It is known that the degree of cation transport is modulated by the presence of ionisable sites on cellulosic paper surfaces (mostly carboxyl groups).29,30,42 A significant amount of carboxyl groups (mmol kg−1 order29,43) is produced by oxidation of cellulosic hydroxyl groups during the industrial paper making process.44 Therefore, the surface of filter paper is negatively charged as demonstrated in previously reported articles.33,45 The electrostatic interaction between a cationic analyte (e.g. Ca2+ in this work) and the anionic paper surface results in the adsorption of target cations onto the filter paper surface.42,46,47 In this adsorption process, the pre-deposited Mg2+ competes with the Ca2+ analyte for the available binding sites on the paper surface. In addition, the adsorption of metal cations, which is described by a Langmuir adsorption isotherm model, is affected by the ionic strength of the sample liquid.48 Since the pre-deposited Mg2+ also contributes to an increase in ionic strength of the flowing liquid, this leads to decreased adsorption of metal cations onto the paper surface. The pre-deposited Mg2+ provides a relatively high but constant ion background, in order to avoid undesired modulation in non-specific Ca2+ adsorption due to sample-induced, non Ca2+-related competition for binding sites or changes in ionic strength. The tolerance against interfering cations relies on the total amount of ionised electrolytes (i.e. pH-buffering reagents, MgCl2 on the sample inlet). For drinking or natural water assays, the WHO guideline addresses the threshold values for multiple ions, including those potentially interfering with the current Ca2+ assay. The values of taste or health-based thresholds are all well below the amounts that can be tolerated by the developed μPAD1 and are therefore expected not to interfere with the distance-based Ca2+ assay.
Sample | Distance-based μPADsa (mmol L−1) | Complexometric titrationb (mmol L−1) | Errorc (%) |
---|---|---|---|
a Each data has been obtained by measurements with 4 individual single-use devices. b Each data has been obtained by 3 individual titrations. c Error estimated by comparing with the experimental results obtained by complexometric titration as reference values. | |||
Mineral water | |||
#1 | 0.43 ± 0.03 | 0.43 ± 0.01 | 0 |
#2 | 0.25 ± 0.03 | 0.23 ± 0.01 | +6 |
#3 | 0.21 ± 0.02 | 0.21 ± 0.01 | +2 |
#4 | 1.96 ± 0.13 | 2.05 ± 0.01 | −5 |
Tap water | |||
#1 | 0.46 ± 0.03 | 0.41 ± 0.02 | +10 |
#2 | 0.48 ± 0.03 | 0.45 ± 0.01 | +6 |
#3 | 0.51 ± 0.03 | 0.44 ± 0.01 | +15 |
#4 | 0.39 ± 0.01 | 0.35 ± 0.01 | +12 |
To further investigate the capability of the distance-based μPADs, quantification of Ca2+ levels in spiked tap water samples was carried out by multiple independent observers. The as-collected tap water (tap water #4 shown in Table 2) as well as 3 spiked samples were applied onto μPADs and the resulting length of the colour-changed detection channel was measured by a ruler with sample exposed μPADs presented to observers in a random order. Before testing, the volunteers were instructed on the readout method using one example of a colour-changed μPAD. Fig. 5 shows the visually read out results for various concentrations of Ca2+ in tap water or spiked tap water samples. Due to the subjective interpretation of the distance-based signals, some differences between individual observers are unavoidable. Overall, the visually detected distance-based signals were within +23% to −11% of the ImageJ software-assisted readout results (Table S2†).
Footnote |
† Electronic supplementary information (ESI) available: Nano-optode preparation, wax pattern design, procedure for obtaining colorimetric response profiles, complexometric titration of water samples, optimisation of printing cycles for nano-optodes, wax barrier tolerance against leakage, optimisation of printing cycles for pH-buffering reagents and MgCl2, comparison between different batches of μPADs, comparison between software-assisted readout and readouts by untrained users, quantification of Ca2+ with commercial paper dipsticks, selectivity evaluation, method comparison of distance-based μPADs with complexometric titration, and evaluation of Ca2+ adsorption to paper substrate and error of observer tests. See DOI: 10.1039/c8an02146e |
This journal is © The Royal Society of Chemistry 2019 |