Analysis of primer gunshot residue particles by laser induced breakdown spectroscopy and laser ablation inductively coupled plasma mass spectrometry

Abstract

This study reports novel approaches for the detection of gunshot residues (GSR) from the hands of individuals using Laser-Induced Breakdown Spectroscopy (LIBS) and Laser Ablation Inductively Coupled Plasma Mass Spectrometry (LA-ICP-MS). The methods’ performance was evaluated using 159 GSR standard and authentic samples. Forty specimens generated from characterized microparticles were used as matrix-matched primer gunshot residue (pGSR) standards to examine the elemental profiles of leaded and lead-free residues, compared to SEM-EDS and solution-ICP-MS. Also, 119 authentic skin samples were analyzed to estimate error rates. Shooter samples were correctly classified into three categories based on their elemental composition (leaded, lead-free, or mixed pGSR). A total of 60 non-shooter samples were used to establish background thresholds and estimate specificity (93.4% for LA-ICP-MS and 100% for LIBS). All the authentic leaded items resulted in the detection of particle(s) with composition characteristic of pGSR (Pb–Ba–Sb), as observed by simultaneous elemental identification of target analytes at the exact ablation times and locations. When considering the pre-characterized elemental composition of these primers as the “ground truth”, LA-ICP-MS resulted in 91.8% sensitivity (true positive rate), while LIBS resulted in 89.2% sensitivity. Particles containing Ba, Bi, Bi–Cu–K, and Cu–Ti–Zn were found in the lead-free residues. Identification of lead-free GSR proved more challenging as some of these elements are common in the environment, resulting in 85.2% sensitivity for LA-ICP-MS and 44.4% for LIBS. Overall accuracies of 94.9% and 88.2% were obtained for the LA-ICP-MS and LIBS sets, respectively. LA-ICP-MS provided an additional level of confidence in the results by its superior analytical capabilities, complementing the LIBS chemical profiles. The laser-based methods provide rapid chemical profiling and micro-spatial information of gunshot residue particles, with minimal destruction of the sample and high accuracy. Chemical mapping of 25 micro-regions per sample is possible in 2–10 minutes by LIBS and LA-ICP-MS, offering new tools for more comprehensive forensic case management and quick GSR screening in environmental and occupational sciences.

Graphical abstract: Analysis of primer gunshot residue particles by laser induced breakdown spectroscopy and laser ablation inductively coupled plasma mass spectrometry

Supplementary files

Article information

Article type
Paper
Submitted
20 Apr 2021
Accepted
29 Jul 2021
First published
03 Aug 2021

Analyst, 2021,146, 5389-5402

Analysis of primer gunshot residue particles by laser induced breakdown spectroscopy and laser ablation inductively coupled plasma mass spectrometry

C. Vander Pyl, C. Martinez-Lopez, K. Menking Hoggatt and T. Trejos, Analyst, 2021, 146, 5389 DOI: 10.1039/D1AN00689D

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.

Social activity

Spotlight

Advertisements