The influence of the laser ablation process on isotopic fractionation of copper in LA-MC-ICP-MS

Abstract

Isotopic fractionation during in-situ analysis of copper using laser ablation (LA) MC-ICP-MS has been investigated. Results show that particles <200 nm transported in the aerosol are isotopically lighter by up to 0.7‰ compared to the larger particles. The dependence of this isotopic fractionation on the ablation time, together with isotopic effects caused in the ICP during particle vaporisation, explains the variations of isotope ratios observed during LA-MC-ICP-MS. LA-MC-ICP-MS measurements of copper isotope ratios in metallic copper typically show large systematic within-run variations of up to several permil (‰) depending on the ablation time and, correspondingly, large offsets from solution data for the same sample when compared directly. Analysis of small particles generated by vapour condensation after the laser impact, and larger particles formed by ejection of liquid droplets from the crater, show that the laser impact results in isotopic fractionation of 65Cu+/63Cu+ between different aerosol particle size fractions of up to 0.7‰. This fractionation cannot be overcome by removal of large particles from the aerosol that are not completely vaporised and ionised in the ICP source. However, separation of the large particles is shown to significantly reduce within-run fractionation, as well as improving the accuracy relative to solution data, and is therefore recommended for the analysis of copper metal. Temporal variations in the isotopic composition of particulate vapour condensate are explained by a crater effect that influences the residence time of large particles in the laser-generated plasma plume, leading to preferential vaporisation of the lighter isotopes.

Graphical abstract: The influence of the laser ablation process on isotopic fractionation of copper in LA-MC-ICP-MS

Article information

Article type
Paper
Submitted
07 Nov 2006
Accepted
27 Feb 2007
First published
09 Mar 2007

J. Anal. At. Spectrom., 2007,22, 547-552

The influence of the laser ablation process on isotopic fractionation of copper in LA-MC-ICP-MS

H. Kuhn, N. J. Pearson and S. E. Jackson, J. Anal. At. Spectrom., 2007, 22, 547 DOI: 10.1039/B616232K

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.

Social activity

Spotlight

Advertisements