Issue 1, 2003

Measurement of density and chemical concentration using a microfluidic chip

Abstract

A new microfluidic product for measuring fluid density, specific gravity and chemical concentration has been developed. At the core of this lab-on-a-chip sensor is a vacuum-sealed resonating silicon microtube. Measurements can be made with under a microliter of sample fluid, which is over 1000× less than is conventionally required. Since the product is MEMS-based the overall system size is a fraction of conventional density meters and it weighs much less than the traditional desk-top, temperature controlled, density meters. The syringe or pipette loaded system includes a dynamic temperature control system that operates between 0 °C and 90 °C with an accuracy of less than 0.01 °C. Density measurement accuracies of 4 to 5 digits have been observed with aqueous solutions. Measurement examples and applications will be discussed.

Article information

Article type
Paper
Submitted
20 Nov 2002
Accepted
10 Jan 2003
First published
03 Feb 2003

Lab Chip, 2003,3, 19-21

Measurement of density and chemical concentration using a microfluidic chip

D. Sparks, R. Smith, M. Straayer, J. Cripe, R. Schneider, A. Chimbayo, S. Anasari and N. Najafi, Lab Chip, 2003, 3, 19 DOI: 10.1039/B211429A

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.