Comparison of two ICP-MS set-ups for measuring 99Tc in large volume water samples

Abstract

Large volume fjord and seawater samples have been radiochemically prepared for ICP-MS analysis in order to test the robustness of the procedure and to carry out a comparison of two ICP-MS set-ups. A sector field instrument (MicroMass PT2) coupled with an ultrasonic nebuliser and a quadrupole ICP-MS (Perkin-Elmer Elan 6000) coupled with an electrothermal vaporisation (ETV) unit were used. The results showed that the radiochemical procedure was robust, removing Ru and Mo to acceptable levels, and that the two set-ups gave results that were in agreement. The correlation coefficient between the sets of 11 results was 1.0 ± 0.05. The importance of establishing the matrix effect when using an ETV is discussed.

Article information

Article type
Paper
Submitted
29 Jun 2001
Accepted
11 Oct 2001
First published
28 Nov 2001

Analyst, 2002,127, 70-75

Comparison of two ICP-MS set-ups for measuring 99Tc in large volume water samples

M. J. Keith-Roach, S. Stürup, D. H. Oughton and H. Dahlgaard, Analyst, 2002, 127, 70 DOI: 10.1039/B105708C

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.

Social activity

Spotlight

Advertisements