Microwave digestion of plant and grain standard reference materials in nitric and hydrofluoric acids for multi-elemental determination by inductively coupled plasma mass spectrometry

Abstract

A microwave-assisted HNO 3 -HF digestion system was explored for the total dissolution of biological plant and grain materials followed by multi-elemental determination using ICP-MS, in order to improve the low recoveries of several elements observed in a previous study using a microwave-assisted nitric acid digestion system. NIST standard reference materials (SRMs), including Apple Leaves (1515), Peach Leaves (1547), Wheat Flour (1567a), Rice Flour (1568a), Tomato Leaves (1573) and Pine Needles (1575), were analyzed. Approximately 0.5 g of sample was digested in 5 ml of HNO 3 and 0.1 ml of HF, with or without a subsequent digestion stage with boric acid. The matrix effect for boron was evaluated for an ICP-MS system and signal enhancement was observed for all the elements tested. Potential spectral interferences in ICP-MS with HNO 3 -HF, boron and biological matrices are discussed and the spectral interferences on Co, As and Se are tabulated. The ICP-MS system was calibrated using external standards prepared in undigested reagent blanks with In as an internal standard. It was found that with a low but sufficient amount of HF in the digestion, the possible precipitation of metal fluorides in the digestate (without boric acid) was not significant. The recoveries for some silicon-bound elements, such as Al, Co, Cr, Ni, Th, U and V, were significantly improved compared with those from digestions with HNO 3 alone. Using the HNO 3 -HF digestion procedure, the ICP-MS results for 30 elements, Al, As, B, Ba, Ca, Cd, Co, Cr, Cu, Fe, Hg, K, Mg, Mn, Mo, Na, Ni, P, Pb, S, Sb, Se, Si, Sr, Th, Ti, Tl, U, V and Zn, agreed well with the certified values in leaf and grain SRMs. The recoveries were mostly within the range 85-115%. Hence, the use of boric acid in the digestion was not necessary, which simplified the procedure, minimized the content of the total dissolved solids in solution for ICP-MS analysis and allowed the determination of boron.

Article information

Article type
Paper

J. Anal. At. Spectrom., 1999,14, 939-946

Microwave digestion of plant and grain standard reference materials in nitric and hydrofluoric acids for multi-elemental determination by inductively coupled plasma mass spectrometry

X. Feng, S. Wu, A. Wharmby and A. Wittmeier, J. Anal. At. Spectrom., 1999, 14, 939 DOI: 10.1039/A804683B

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.

Social activity

Spotlight

Advertisements