FIA method for the determination of vitamin A by iodine

Abstract

An FIA method for the determination of vitamin A in an aqueous solution is described. A two-line manifold was used with an iodine solution as the reagent stream and water as the carrier stream. Iodine was reduced by retinol to produce the iodide ion, the absorbance band of which was measured at 226 nm. The conditions for the FIA system were optimized. A calibration curve was obtained, which was linear between 0.1–50 and 50–100 mg l–1. The correlation coefficients were 0.999 and 0.998 and the limit of detection was calculated to be 53 µg l–1; the sr values lay between 0.39 and 12.37%.

Article information

Article type
Paper

Anal. Proc., 1995,32, 141-143

FIA method for the determination of vitamin A by iodine

N. Spatny, S. J. Haswell and M. Grasserbauer, Anal. Proc., 1995, 32, 141 DOI: 10.1039/AI9953200141

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.