Issue 8, 1994

Sensitive determination of periodate and tartaric acid by stopped-flow chemiluminescence spectrometry

Abstract

The stopped-flow technique was used for the determination of periodate by measuring the chemiluminescence produced on reaction with luminol, enchanced by manganese(II) and triethanolamine. The method was also applied to the determination of tartaric acid by oxidation with periodate. Under the optimum conditions, responses were linear between 1.0 × 10–6 and 2.5 × 10–4 mol l–1 periodate and from 5.0 × 10–6 to 5.0 × 10–5 mol l–1 tartaric acid, with a relative standard deviation below about 2% in both instances. The proposed method compares favourably with existing alternatives in terms of sensitivity and sampling frequency. It was used for the determination of tartaric acid in pharmaceutical preparations, the results being consistent with the manufacturers' certified contents.

Article information

Article type
Paper

Analyst, 1994,119, 1819-1824

Sensitive determination of periodate and tartaric acid by stopped-flow chemiluminescence spectrometry

A. Gaikwad, M. Silva and D. Pérez-Bendito, Analyst, 1994, 119, 1819 DOI: 10.1039/AN9941901819

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.

Social activity

Spotlight

Advertisements